错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

A Driller-Tipped Guidewire Design with Combined Magnetic Navigation and Drilling Motion

  • Li Zhang,
  • Tiantian Xu,
  • Mingxue Cai,
  • Moqiu Zhang

摘要

The endovascular intervention has been widely used to treat occlusive peripheral vascular disease (PVD). However, the current procedure has concerns such as low successful rate, undesired trauma, and radiation exposure. This chapter proposes a robotic solution to mitigate these limitations. A miniature magnetic driller-tipped guidewire (MDG) is designed, which performs (1) controllable bending under directional magnetic fields to navigate in complex vasculature and (2) mechanical drilling under rotating magnetic fields to pass through clogged regions. An integrated actuation system is adopted for magnetic control and ultrasound (US) imaging in a large workspace. A control framework composed of the preoperative and intraoperative stages is developed, which internally coordinates all system modules, addresses noisy US imaging, and provides convenient operation. Demonstrations under an optical camera verify the flexible steering ability and effective unclogging motion of the MDG. Furthermore, US image-monitored in vitro experiments validate the overall hardware platform and control strategy. The designed MDG realizes dual objectives, including controllability at bifurcations and penetrability at clots. The introduced US imaging modality largely reduces radiation hazards in conventional catheterization. This chapter studies the intervention of the MDG under electromagnetic field navigation and US imaging guidance. The proposed workflow has the potential to improve operational safety and clinical outcomes of the occlusive PVD treatment.