Fabrication of CMUT with T-shaped Cavities Using Anodic Bonding Process

  • Shaohui Qin,
  • Jiawei Yuan,
  • Shiwang Zhang,
  • Zhixuan Li,
  • Yihe Zhao,
  • Zhikang Li,
  • Libo Zhao

摘要

Capacitive micromachined ultrasonic transducers (CMUT) are known for their high sensitivity, favorable impedance matching, broad bandwidth, and good biocompatibility, making them suitable for medical diagnostics, non-destructive testing, and 3D ultrasound gesture recognition. However, challenges such as high collapse voltages (≥100 V) and difficulties in enhancing both transmitting and receiving sensitivities limit their broader application. In this paper, we introduce a novel T-shaped cavity CMUT (T-CMUT) design that utilizes the electrostatic stiffness softening effect. By reducing the electrode spacing in the cavity’s peripheral region, we achieved a higher electric field intensity, resulting in decreased stiffness and improved piston-like deformation of the membrane. Additionally, we developed a fabrication process for T-CMUT using silicon-glass anodic bonding, addressing stringent surface roughness requirements and high preparation temperatures associated with traditional wafer direct bonding processes. The results indicate that the T-CMUT exhibit a 20.6% increase in transmission sensitivity and a 36.6% improvement in fractional bandwidth compared to conventional CMUTs.