Microwave Tomography requires calibration to accurately relate measured S-parameters to electric fields, accounting for antenna imperfections and modeling assumptions. We present an experimental setup using amplitude-only data from a synthetic antenna array immersed in a coupling liquid. Two-dimensional simulations including the container were performed to explain the measurements. Additionally, a calibration procedure is proposed that integrates data with and without two different scatterers.

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Phaseless Microwave Imaging Setup: Comparison of Simulations and Experiments

  • María José Cervantes,
  • Catalina A. Cely-Ortiz,
  • Marisa Orzuza,
  • Julián Galván,
  • Ruben Moran Fabra,
  • Enrique Berjano,
  • César F. Caiafa,
  • Ramiro M. Irastorza

摘要

Microwave Tomography requires calibration to accurately relate measured S-parameters to electric fields, accounting for antenna imperfections and modeling assumptions. We present an experimental setup using amplitude-only data from a synthetic antenna array immersed in a coupling liquid. Two-dimensional simulations including the container were performed to explain the measurements. Additionally, a calibration procedure is proposed that integrates data with and without two different scatterers.