Minimally invasive surgery procedures require the surgical tools to be moved about an incision point. Robots used for these surgeries must achieve this motion capability. When robots are used alongside the surgeon, additional considerations must be made regulating the surgeon-robot interaction. This paper presents a formulation of a compliance controller, namely an admittance controller, for use in the control of a surgical robot that is directed by physical interaction with the surgeon. The surgical robot in this study is a 6 degrees-of-freedom collaborative robot with a force/torque sensor attached to the distal end of its last joint.

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Formulation of Admittance Control of a Surgical Robot About Remote Center of Motion

  • Mehmet Alp Balkan,
  • Mehmet İsmet Can Dede

摘要

Minimally invasive surgery procedures require the surgical tools to be moved about an incision point. Robots used for these surgeries must achieve this motion capability. When robots are used alongside the surgeon, additional considerations must be made regulating the surgeon-robot interaction. This paper presents a formulation of a compliance controller, namely an admittance controller, for use in the control of a surgical robot that is directed by physical interaction with the surgeon. The surgical robot in this study is a 6 degrees-of-freedom collaborative robot with a force/torque sensor attached to the distal end of its last joint.