This work presents an 8-channel neuronal signal acquisition and conditioning system with the ability to drive LED on the optrode made of on-the-shelf discrete components. The system is divided into 3 modules: a DC offset generating module, an LED driving module, and a two-stage signal amplifying and conditioning module. Texas Instruments LMP770x series amplifiers were selected to be the amplifiers of this system. The system is designed to be powered with batteries to minimize electromagnetic interference. A PCB was fabricated to verify the system. The gain of the system is 41.3 dB and the bandwidth of the system is 80 Hz to 14 kHz. To verify the ability of the system in laboratory environment, an in vivo recording of the Inferior Colliculus of laboratory mice was conducted.

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

AN 8-Channel Neuronal Signal Acquisition and Conditioning System with Optrode LED Driving Capability

  • Hong Liang Loo,
  • Xiangxin Ma,
  • Yu Xia,
  • Yujing Huang,
  • Hung Chun Li,
  • Zhen Yuan,
  • Peng Un Mak,
  • Mang I. Vai,
  • Sio Hang Pun

摘要

This work presents an 8-channel neuronal signal acquisition and conditioning system with the ability to drive LED on the optrode made of on-the-shelf discrete components. The system is divided into 3 modules: a DC offset generating module, an LED driving module, and a two-stage signal amplifying and conditioning module. Texas Instruments LMP770x series amplifiers were selected to be the amplifiers of this system. The system is designed to be powered with batteries to minimize electromagnetic interference. A PCB was fabricated to verify the system. The gain of the system is 41.3 dB and the bandwidth of the system is 80 Hz to 14 kHz. To verify the ability of the system in laboratory environment, an in vivo recording of the Inferior Colliculus of laboratory mice was conducted.