Accuracy and Repeatability of a Parallel Robot for Personalised Minimally Invasive Surgery

  • Doina Pisla,
  • Paul Tucan,
  • Damien Chablat,
  • Nadim Al Hajjar,
  • Andra Ciocan,
  • Alexandru Pusca,
  • Adrian Pisla,
  • Corina Radu,
  • Grigore Pop,
  • Bogdan Gherman

摘要

The paper presents the methodology used for accuracy and repeatability measurements of the experimental model of a parallel robot developed for surgical applications. The experimental setup uses a motion tracking system (for accuracy) and a high precision measuring arm for position (for repeatability). The accuracy was obtained by comparing the trajectory data from the experimental measurement with a baseline trajectory defined with the kinematic models of the parallel robotic system. The repeatability was experimentally determined by moving (repeatedly) the robot platform in predefined points.