错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

A Novel Force Sensing Biopsy Needle for MR-Compatible Robotic Biopsy System: Performance Evaluation

  • Rongrong Liu,
  • Seong Young Ko

摘要

A force-sensing biopsy needle embedded with Fiber Bragg Grating (FBG) sensors is proposed to be used in cancer diagnosis. The FBG sensor is a fiber optic sensor, and the biopsy needle is made of nitinol, both of which can be used in MRI-guided biopsy to provide force feedback to surgeons. The axial force resolution of the biopsy needle is as high as 8.4 mN/pm, and its repeatability was measured to be 19 mN. The dynamic force measurement performance of the needle was verified by animal meat insertion experiment.