Study Objectives: <p>Despite its efficacy, long-term adherence to continuous positive airway pressure (CPAP) hovers around 50%. The V-Com (SleepRes, Murfreesboro, TN), an inspiratory pressure relief (IPR) accessory, claims to improve comfort by decreasing inspiratory pressures. We aim to independently assess its impact on fixed CPAP settings in a laboratory bench model.</p> Methods: <p>An ASL-5000 Breathing simulator (IngMar Medical, Pittsburgh, PA) with normal lung settings was employed. A ResMed AirSense 11 device was connected to ASL-5000 through a React Health Rio II exhalation vent and an elbow adapter. Mean inspiratory and expiratory pressures were measured at fixed CPAP settings (4, 6, 8, 10, 12, 14, 16, 18, and 20 cm H<sub>2</sub>O) at baseline, 1 IPR unit, and 2 IPR units in series, without and with expiratory pressure relief. Generalized linear models were applied to evaluate changes in pressure across CPAP settings.</p> Results: <p>With 1 IPR, mean inspiratory pressures decreased significantly from the baseline pressures (<i>P</i> &lt; .001), with only a marginal reduction of mean expiratory pressures (<i>P</i> = .239). The inspiratory pressure reduction was consistently greater than the expiratory pressure drop across all settings (<i>P</i> &lt; .001). Higher CPAP settings resulted in larger reductions in inspiratory and expiratory pressures (<i>P</i> &lt; .001). Two IPR units in series led to a greater decrease in both pressures (<i>P</i> = .001). When expiratory pressure relief 3 was combined with IPR, a larger drop in pressure was noted (<i>P</i> &lt; .001).</p> Conclusions: <p>This IPR accessory decreases inspiratory pressures but only marginally reduces expiratory pressures in this laboratory bench model. Two IPR units further decrease both inspiratory and expiratory pressures.</p> Citation: <p>Sleeper GP, Catullo KA, Chen P-L, Patil SP, Strohl KP, Chiang AA. The influence of an innovative inspiratory pressure relief accessory on continuous positive airway pressure profiles in a laboratory bench setting. <i>J Clin Sleep Med</i>. 2025;21(6):1073–1079.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The influence of an innovative inspiratory pressure relief accessory on continuous positive airway pressure profiles in a laboratory bench setting

  • Geoffery P. Sleeper,
  • Kimber A. Catullo,
  • Pai-Lien Chen,
  • Susheel P. Patil,
  • Kingman P. Strohl,
  • Ambrose A. Chiang

摘要

Study Objectives:

Despite its efficacy, long-term adherence to continuous positive airway pressure (CPAP) hovers around 50%. The V-Com (SleepRes, Murfreesboro, TN), an inspiratory pressure relief (IPR) accessory, claims to improve comfort by decreasing inspiratory pressures. We aim to independently assess its impact on fixed CPAP settings in a laboratory bench model.

Methods:

An ASL-5000 Breathing simulator (IngMar Medical, Pittsburgh, PA) with normal lung settings was employed. A ResMed AirSense 11 device was connected to ASL-5000 through a React Health Rio II exhalation vent and an elbow adapter. Mean inspiratory and expiratory pressures were measured at fixed CPAP settings (4, 6, 8, 10, 12, 14, 16, 18, and 20 cm H2O) at baseline, 1 IPR unit, and 2 IPR units in series, without and with expiratory pressure relief. Generalized linear models were applied to evaluate changes in pressure across CPAP settings.

Results:

With 1 IPR, mean inspiratory pressures decreased significantly from the baseline pressures (P < .001), with only a marginal reduction of mean expiratory pressures (P = .239). The inspiratory pressure reduction was consistently greater than the expiratory pressure drop across all settings (P < .001). Higher CPAP settings resulted in larger reductions in inspiratory and expiratory pressures (P < .001). Two IPR units in series led to a greater decrease in both pressures (P = .001). When expiratory pressure relief 3 was combined with IPR, a larger drop in pressure was noted (P < .001).

Conclusions:

This IPR accessory decreases inspiratory pressures but only marginally reduces expiratory pressures in this laboratory bench model. Two IPR units further decrease both inspiratory and expiratory pressures.

Citation:

Sleeper GP, Catullo KA, Chen P-L, Patil SP, Strohl KP, Chiang AA. The influence of an innovative inspiratory pressure relief accessory on continuous positive airway pressure profiles in a laboratory bench setting. J Clin Sleep Med. 2025;21(6):1073–1079.