A Monte Carlo Simulation Study of the Optimal Source Offset for Scatter Fraction Estimation in PET and the Influence of the Scanner and Object Characteristics
摘要
In positron emission tomography (PET), the scattered photons represent a major image degrading factor. The scatter fraction (SF) of PET scanners is experimentally determined following the National Electrical Manufacturers Association (NEMA) NU 2 protocol, using a polyethylene phantom with a line source inserted at 45 mm radial offset from the phantom’s center. In this work, we aim to investigate the optimal radial offset of the line source, that represents the SF of a uniformly activated phantom, for the lutetium yttrium orthosilicate (LYSO)-based uEXPLORER total body PET scanner. Additionally, we propose to study the impact of several factors (axial field-of-view (AFOV), crystal material and heterogeneities) on this optimal offset, in an effort to provide recommendations for a wide range of PET scanner configurations. We performed Monte Carlo (MC) simulations of the uEXPLORER scanner along with five phantoms of different diameters, using the Geant4 Application for Tomographic Emission (GATE) code. We performed supplemental MC simulations in which we varied the crystal material, the AFOV and we introduced bone and lung heterogeneities in the phantoms to assess their impact on the optimal offset findings. Our results demonstrated that the 45 mm offset overestimates the SF of the uniform 20 cm diameter phantom by 24.3