Simple surgical explant technique for the EDWARDS INTUITY rapid deployment valve: a case report of prosthetic valve endocarditis
摘要
A rapid deployment valve can shorten operation times and improve hemodynamics. However, explantation can be challenging because of the unique structure of such valves, including an inflow frame covered by textured sealing cloth beneath the sewing cuff. In this case, we report a simple explantation technique.
Case presentationThis case involved a 79-year-old woman with prosthetic valve endocarditis who had undergone aortic valve replacement with a rapid deployment valve 5 years earlier. Preoperative echocardiography revealed severe mitral regurgitation and a highly mobile mass on the posterior leaflet. The prosthetic valve had thickened cusps without regurgitation. Emergent surgery was performed to explant the prosthetic valve and replace both the aortic and mitral valves through a re-median sternotomy under routine cardiopulmonary bypass support. The textured sealing cuff was detached from the surrounding tissue after separating the rigid outflow portion from the transformable inflow portion by cutting the fabric. No annular or sub-annular damage was observed. Enterococcus faecalis was cultured from the blood. The patient received 6 weeks of antimicrobial therapy and was discharged without symptoms of heart failure or infection.
ConclusionThe patient successfully underwent valve explantation and double valve replacement for prosthetic valve endocarditis. This method is safe and feasible for explanting rapid deployment valves with minimal tissue damage.