Background <p>The concurrent measurement of emtricitabine, tenofovir, and rilpivirine was achieved using UPLC, employing Analytical Quality by Design (AQbD) principles, which encompass response surface methodology (RSM) and Box–Behnken design (BBD). This approach offers a systematic and efficient method for improving analytical procedures and ensuring reliable results.</p> Results <p>Analyte separation was accomplished using an Endoversil C 18 (2.1 × 50&#xa0;mm, 1.7&#xa0;µm) column as the stationary phase. The mobile phase consisted of a mixture of 80% orthophosphoric acid (0.1%) and 20% acetonitrile, with a flow rate of 0.4&#xa0;ml/min and a column temperature of 30&#xa0;°C. A photodiode array detector was used for detection at 270&#xa0;nm. The peak area response-concentration curve exhibited linearity across ranges of 50–250&#xa0;µg/ml (emtricitabine), 100–500&#xa0;µg/ml (tenofovir), and 10–50&#xa0;µg/ml (rilpivirine). Quantitation limits were determined to be 0.099&#xa0;µg/ml (emtricitabine), 0.165&#xa0;µg/ml (tenofovir), and 0.066&#xa0;µg/ml (rilpivirine). The method was successfully validated for the simultaneous determination of emtricitabine, tenofovir, and rilpivirine in combined tablet dosage form. Percentage recoveries were 99.89%, 99.52%, and 100.20%, with relative standard deviations of 0.415%, 0.268%, and 0.559% for emtricitabine, tenofovir, and rilpivirine, respectively.</p> Conclusion <p>The performance of the proposed method was evaluated against reported RP-UPLC methods and found to be rapid and cost-effective. The developed and validated stability indicating RP-UPLC method proved suitable for quality control and drug analysis.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Stability indicating RP-UPLC determination of three antiviral agents: emtricitabine, tenofovir, and rilpivirine in combined pharmaceutical dosage form

  • Vinutha Kommineni,
  • Spoorthi Pohar,
  • N. Sri Lakshmi,
  • N. Swarna Latha

摘要

Background

The concurrent measurement of emtricitabine, tenofovir, and rilpivirine was achieved using UPLC, employing Analytical Quality by Design (AQbD) principles, which encompass response surface methodology (RSM) and Box–Behnken design (BBD). This approach offers a systematic and efficient method for improving analytical procedures and ensuring reliable results.

Results

Analyte separation was accomplished using an Endoversil C 18 (2.1 × 50 mm, 1.7 µm) column as the stationary phase. The mobile phase consisted of a mixture of 80% orthophosphoric acid (0.1%) and 20% acetonitrile, with a flow rate of 0.4 ml/min and a column temperature of 30 °C. A photodiode array detector was used for detection at 270 nm. The peak area response-concentration curve exhibited linearity across ranges of 50–250 µg/ml (emtricitabine), 100–500 µg/ml (tenofovir), and 10–50 µg/ml (rilpivirine). Quantitation limits were determined to be 0.099 µg/ml (emtricitabine), 0.165 µg/ml (tenofovir), and 0.066 µg/ml (rilpivirine). The method was successfully validated for the simultaneous determination of emtricitabine, tenofovir, and rilpivirine in combined tablet dosage form. Percentage recoveries were 99.89%, 99.52%, and 100.20%, with relative standard deviations of 0.415%, 0.268%, and 0.559% for emtricitabine, tenofovir, and rilpivirine, respectively.

Conclusion

The performance of the proposed method was evaluated against reported RP-UPLC methods and found to be rapid and cost-effective. The developed and validated stability indicating RP-UPLC method proved suitable for quality control and drug analysis.