Stem anteversion is not affected by proximal femur geometry in robotic-assisted total hip arthroplasty

  • Andrea Marcovigi,
  • Gianluca Grandi,
  • Luca Bianchi,
  • Francesco Zambianchi,
  • Marco Pavesi,
  • Fabio Catani

摘要

Background

In the present study, the surgeon aimed to align the stem at 5° to 25° in anteversion. The robotic technology was used to measure stem anteversion with respect to proximal femur anteversion at different levels down the femur.

Methods

A total of 102 consecutive patients underwent robotic-arm-assisted total hip arthroplasty (RTHA). 3D CT-based preoperative planning was performed to determine femoral neck version (FNV), posterior cortex anteversion (PCA), anterior cortex anteversion (ACA), and femoral metaphyseal axis anteversion (MAA) at 3 different levels: D (10 mm above lesser trochanter), E (the midpoint of the planned neck resection line) and F (head-neck junction). The robotic system was used to define and measure stem anteversion during surgery.

Results

Mean FNV was 6.6° (SD: 8.8°) and the mean MAA was consistently significantly higher than FNV, growing progressively from proximal to distal. Mean SV was 16.4° (SD: 4.7°). There was no statistically significant difference (P = 0.16) between SV and MAA at the most distal measured level. In 96.1% cases, the stem was positioned inside the 5°–25° anteversion range.

Conclusions

Femoral anteversion progressively increased from neck to proximal metaphysis. Aligning the stem close to femoral anteversion 10 mm above the lesser trochanter often led to the desired component anteversion.