Background <p>We compared ultrahigh-resolution (UHR) photon-counting detector-computed tomography (PCD-CT) and spectral post-processed images for coronary stent visualization in a dynamic, anthropomorphic, and circulatory phantom.</p> Materials and methods <p>Ten coronary stents were scanned at 60, 80, and 100 beats per min (bpm) using UHR-spectral PCD-CT (96 × 0.2 mm collimation). Reconstructions included UHR (0.2 mm), downsampled (0.6 mm), and spectral&#xa0;post-processed images (0.4 mm), including virtual monoenergetic images (VMI; 45–100 keV), lumen-preserving images, and iodine maps (IM). Objective quality was assessed by measurable&#xa0;stent lumen visibility and stent&#xa0;strut width overestimation&#xa0;factor,&#xa0;compared&#xa0;to nominal strut width. Subjective quality was rated using a 4-point Likert scale. Repeated-measures analysis of variance−ANOVA and Friedman test with <i>post hoc</i> corrections were applied.</p> Results <p>UHR images provided the highest lumen visibility (62.6 ± 7.6%) at 60 bpm, outperforming all reconstructions ranging 43.5–52.7% (<i>p</i> ≤ 0.001) except IM (59.6 ± 11.9%, pairwise&#xa0;<i>p</i> = 0.839). UHR showed the lowest strut overestimation factor&#xa0;(18.1 ± 3.4), better than&#xa0;all spectral images (20.2–26.2, <i>p</i> ≤ 0.003) and DS (32.1 ± 6.5, <i>p</i> &lt; 0.001). Subjective quality was best for UHR at 60 bpm (4.0 [interquartile range, IQR 4.0–4.0]) but declined at 100 bpm (3.0 [IQR 2.0–3.0], <i>p</i> &lt; 0.01). VMI at 55 keV and IM maintained stable quality across heart rates (<i>p</i> ≥ 0.09).</p> Conclusion <p>PCD-CT combining UHR and spectral imaging enhances stent assessment. UHR provides the best lumen visibility and strut accuracy but suffers from motion artifacts, whereas VMIs at 55 keV and IM remain stable across heart rates and potentially provide incremental value.</p> Relevance statement <p>Combining UHR and spectral PCD-CT enhances coronary stent visualization by balancing high spatial detail with artifact reduction, potentially improving diagnostic confidence and enabling more reliable non-invasive follow-up across a range of heart rates in clinical practice.</p> Key Points <p><UnorderedList Mark="Bullet"> <ItemContent> <p>The combined value of spectral UHR CT for stent imaging remains largely unexplored.</p> </ItemContent> <ItemContent> <p>UHR PCD-CT showed the highest lumen visibility and sharpest strut delineation, whilst being prone to motion artifacts.</p> </ItemContent> <ItemContent> <p>Spectral reconstructions complement UHR by reducing artifacts and stabilizing image quality, especially at higher heart rates.</p> </ItemContent> </UnorderedList></p> Graphical Abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Spectral ultrahigh-resolution photon-counting CT for coronary stent imaging: evaluation in a dynamic phantom

  • Muhammad Taha Hagar,
  • Tilman Emrich,
  • Milán Vecsey-Nagy,
  • Fabian Bamberg,
  • Christopher L. Schlett,
  • G. William Garrison,
  • Ava Wenderoth,
  • Alexander Isaak,
  • Daniel Kuetting,
  • Julian A. Luetkens,
  • Constantin von Zur Mühlen,
  • Akos Varga-Szemes,
  • Dmitrij Kravchenko

摘要

Background

We compared ultrahigh-resolution (UHR) photon-counting detector-computed tomography (PCD-CT) and spectral post-processed images for coronary stent visualization in a dynamic, anthropomorphic, and circulatory phantom.

Materials and methods

Ten coronary stents were scanned at 60, 80, and 100 beats per min (bpm) using UHR-spectral PCD-CT (96 × 0.2 mm collimation). Reconstructions included UHR (0.2 mm), downsampled (0.6 mm), and spectral post-processed images (0.4 mm), including virtual monoenergetic images (VMI; 45–100 keV), lumen-preserving images, and iodine maps (IM). Objective quality was assessed by measurable stent lumen visibility and stent strut width overestimation factor, compared to nominal strut width. Subjective quality was rated using a 4-point Likert scale. Repeated-measures analysis of variance−ANOVA and Friedman test with post hoc corrections were applied.

Results

UHR images provided the highest lumen visibility (62.6 ± 7.6%) at 60 bpm, outperforming all reconstructions ranging 43.5–52.7% (p ≤ 0.001) except IM (59.6 ± 11.9%, pairwise p = 0.839). UHR showed the lowest strut overestimation factor (18.1 ± 3.4), better than all spectral images (20.2–26.2, p ≤ 0.003) and DS (32.1 ± 6.5, p < 0.001). Subjective quality was best for UHR at 60 bpm (4.0 [interquartile range, IQR 4.0–4.0]) but declined at 100 bpm (3.0 [IQR 2.0–3.0], p < 0.01). VMI at 55 keV and IM maintained stable quality across heart rates (p ≥ 0.09).

Conclusion

PCD-CT combining UHR and spectral imaging enhances stent assessment. UHR provides the best lumen visibility and strut accuracy but suffers from motion artifacts, whereas VMIs at 55 keV and IM remain stable across heart rates and potentially provide incremental value.

Relevance statement

Combining UHR and spectral PCD-CT enhances coronary stent visualization by balancing high spatial detail with artifact reduction, potentially improving diagnostic confidence and enabling more reliable non-invasive follow-up across a range of heart rates in clinical practice.

Key Points

The combined value of spectral UHR CT for stent imaging remains largely unexplored.

UHR PCD-CT showed the highest lumen visibility and sharpest strut delineation, whilst being prone to motion artifacts.

Spectral reconstructions complement UHR by reducing artifacts and stabilizing image quality, especially at higher heart rates.