Background <p>Catheters used for magnetic resonance (MR)-guided interventions require intra-catheter coils and often produce artifacts. This study aimed to fabricate 3D-printed catheters impregnated with vitamin D solution to allow for optimal visualization during MR-guided procedures.</p> Methods <p>3D printing was used to fabricate catheters impregnated with vitamin D solution. Computer-aided design files were generated for a size 18 French catheter prototype with a compartment for vitamin D solution to be manually introduced into the catheter’s lumen and sealed via thermoplastic welding. Polylactic acid (PLA) bioplastic was 3D printed into filaments via material extrusion (FDM<sup>®</sup>, Stratasys, Eden Prairie, MN) on a 5th generation Replicator 3D printer (MakerBot). Three different forms of vitamin D were used, cholecalciferol, ergocalciferol, and calcitriol, and 0.9% normal saline served as a control. Three prints of each catheter type were fabricated and scanned using a 1.5 T MR whole body scanner (Avanto, Siemens Healthcare) inside a small flex loop surface radiofrequency (RF) coil. A 3D gradient recalled echo (GRE) sequence was used with the following acquisition parameters: 4.52/11 ms TE/TR, 15° flip angle, 256 × 256 matrix with 0.5&#xa0;mm × 0.5&#xa0;mm in-plane resolution, 24 coronal slabs, 2&#xa0;mm thickness, and 140&#xa0;Hz receiver bandwidth. Three averages were used to improve the signal-to-noise ratio (SNR). The GRE sequence was run with 4 different flip angles: 3°, 15°, 30°, and 45° to perform T1 mapping.</p> Results <p>All 3D-printed catheters impregnated with vitamin D produced a signal on MR. SNR for vitamin D catheters was similar across the various forms of vitamin D: mean SNRs for 100% cholecalciferol, ergocalciferol, and calcitriol were 138, 139, and 130. Mean SNR and contrast-to-noise ratio (CNR) for vitamin D catheters were significantly higher than the control saline catheter (<i>p</i> &lt; 0.001, for both SNR and CNR). T1 values were lower in vitamin D-impregnated catheters compared to the saline control (228 ± 67 ms and 3371 ± 493 ms, respectively; <i>p</i> &lt; 0.0001), indicating a better signal.</p> Conclusions <p>3D printing of catheters impregnated with vitamin D is feasible and can potentially optimize MR-guided procedures.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

3D printed vitamin D impregnated catheters for magnetic resonance-guided interventions: proof of concept and imaging characteristics

  • Liam O. Cunningham,
  • Aravinda Ganapathy,
  • Cihat Eldeniz,
  • Jeffery A. Weisman,
  • Kevin E. Lindsay,
  • Udayabhanu Jammalamadaka,
  • Karthik Tappa,
  • Amber Salter,
  • Hongyu An,
  • Pamela K. Woodard,
  • David H. Ballard

摘要

Background

Catheters used for magnetic resonance (MR)-guided interventions require intra-catheter coils and often produce artifacts. This study aimed to fabricate 3D-printed catheters impregnated with vitamin D solution to allow for optimal visualization during MR-guided procedures.

Methods

3D printing was used to fabricate catheters impregnated with vitamin D solution. Computer-aided design files were generated for a size 18 French catheter prototype with a compartment for vitamin D solution to be manually introduced into the catheter’s lumen and sealed via thermoplastic welding. Polylactic acid (PLA) bioplastic was 3D printed into filaments via material extrusion (FDM®, Stratasys, Eden Prairie, MN) on a 5th generation Replicator 3D printer (MakerBot). Three different forms of vitamin D were used, cholecalciferol, ergocalciferol, and calcitriol, and 0.9% normal saline served as a control. Three prints of each catheter type were fabricated and scanned using a 1.5 T MR whole body scanner (Avanto, Siemens Healthcare) inside a small flex loop surface radiofrequency (RF) coil. A 3D gradient recalled echo (GRE) sequence was used with the following acquisition parameters: 4.52/11 ms TE/TR, 15° flip angle, 256 × 256 matrix with 0.5 mm × 0.5 mm in-plane resolution, 24 coronal slabs, 2 mm thickness, and 140 Hz receiver bandwidth. Three averages were used to improve the signal-to-noise ratio (SNR). The GRE sequence was run with 4 different flip angles: 3°, 15°, 30°, and 45° to perform T1 mapping.

Results

All 3D-printed catheters impregnated with vitamin D produced a signal on MR. SNR for vitamin D catheters was similar across the various forms of vitamin D: mean SNRs for 100% cholecalciferol, ergocalciferol, and calcitriol were 138, 139, and 130. Mean SNR and contrast-to-noise ratio (CNR) for vitamin D catheters were significantly higher than the control saline catheter (p < 0.001, for both SNR and CNR). T1 values were lower in vitamin D-impregnated catheters compared to the saline control (228 ± 67 ms and 3371 ± 493 ms, respectively; p < 0.0001), indicating a better signal.

Conclusions

3D printing of catheters impregnated with vitamin D is feasible and can potentially optimize MR-guided procedures.