Background <p>Recurrent craniopharyngiomas pose high risks of postoperative visual dysfunction (POVD) during surgery. The current study aimed to explore the application value of intraoperative visual evoked potential (VEP) monitoring during the extended endonasal endoscopic approach (EEEA) for recurrent craniopharyngiomas.</p> Methods <p>A total of 42 patients with recurrent craniopharyngiomas undergoing EEEA with VEP monitoring were analyzed. The amplitude reduction ratios of N75-P100 and P100-N145 were calculated, and their predictive values for POVD were evaluated using group comparisons, receiver operating characteristic (ROC) curve analysis, and binary logistic regression analysis.</p> Results <p>POVD was observed in 8 eyes (8/84, 9.52%) from 7 patients (7/42, 16.67%). Eyes with POVD exhibited significantly greater N75-P100 and P100-N145 amplitude reduction ratios than those without (p &lt; 0.001 and p = 0.002, respectively). The threshold values of the two ratios for predicting POVD were 36.59% (AUC 0.862, p &lt; 0.001) and 36.65% (AUC 0.791, p=0.007), respectively. Multivariate analysis identified that abnormal N75-P100 change was the sole independent predictor of POVD (Odds ratio 9.257, 95% Confidence interval 1.124-76.263; p = 0.039).</p> Conclusions <p>Intraoperative VEP monitoring was particularly recommended for patients undergoing EEEA for recurrent craniopharyngiomas. A one-third reduction in N75-P100 amplitude was proposed as an early warning criterion for VEP monitoring in this patient population.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The predictive value of intraoperative visual evoked potential monitoring for postoperative visual outcomes following extended endoscopic endonasal resection of recurrent craniopharyngiomas

  • Xiaorong Tao,
  • Ke Li,
  • Xiaocui Yang,
  • Jiajia Liu,
  • Jun Yang,
  • Jiawei Shi,
  • Yingzhun Liang,
  • Songbai Gui,
  • Chuzhong Li,
  • Xing Fan,
  • Hui Qiao

摘要

Background

Recurrent craniopharyngiomas pose high risks of postoperative visual dysfunction (POVD) during surgery. The current study aimed to explore the application value of intraoperative visual evoked potential (VEP) monitoring during the extended endonasal endoscopic approach (EEEA) for recurrent craniopharyngiomas.

Methods

A total of 42 patients with recurrent craniopharyngiomas undergoing EEEA with VEP monitoring were analyzed. The amplitude reduction ratios of N75-P100 and P100-N145 were calculated, and their predictive values for POVD were evaluated using group comparisons, receiver operating characteristic (ROC) curve analysis, and binary logistic regression analysis.

Results

POVD was observed in 8 eyes (8/84, 9.52%) from 7 patients (7/42, 16.67%). Eyes with POVD exhibited significantly greater N75-P100 and P100-N145 amplitude reduction ratios than those without (p < 0.001 and p = 0.002, respectively). The threshold values of the two ratios for predicting POVD were 36.59% (AUC 0.862, p < 0.001) and 36.65% (AUC 0.791, p=0.007), respectively. Multivariate analysis identified that abnormal N75-P100 change was the sole independent predictor of POVD (Odds ratio 9.257, 95% Confidence interval 1.124-76.263; p = 0.039).

Conclusions

Intraoperative VEP monitoring was particularly recommended for patients undergoing EEEA for recurrent craniopharyngiomas. A one-third reduction in N75-P100 amplitude was proposed as an early warning criterion for VEP monitoring in this patient population.