Comparison of human milk bottle with infant/toddler test weights in the community setting

  • Anita L. Esquerra-Zwiers,
  • Madeline Heter,
  • Anastasia Perecki,
  • Olivia Jackson,
  • D. Addam Jongekryg,
  • Brian Yurk

摘要

Background

The accuracy of infant intake using test weights (TWs), the change in weight before and after an infant feeds, has only been validated in hospitalized premature infants. This study’s primary aim was to identify how accurate parent infants/toddler (< 2 years old) TWs are at measuring infant intake.

Methods

Data were collected from 101 paired bottle and infant/toddler TWs with 31 participants. Parents participated in the feeding sessions by completing infant/toddler TWs blinded to the researcher. Research assistants completed human milk bottle TWs. Infants were fed previously expressed human milk, initially 30 g, but volumes were increased to not exceed the scale’s capacity.

Results

The mean difference between the bottle TWs measured using the Tanita and OHAUS scales was not significantly different from zero (95% CI (Tanita – OHAUS): (-0.251, 0.108) g). The mean difference between infant/toddler and bottle TWs was significantly different from 0 (95% CI (infant—bottle): (-3.45, -0.915 g or -3.57, -0.95 mL). Infant/toddler and bottle TWs were in agreement with a difference of 2.18 g (SD = 6.63) or 2.25 mL within the scale stated accuracy.

Conclusions

The Tanita infant digital scale accurately measures bottle TWs. The differences in parent infant/toddler TWs are within a clinically acceptable range.