Purpose <p>While data-driven motion correction (DDMC) techniques have proven to enhance the visibility of lesions affected by motion, their impact on overall detectability remains unclear. This study investigates whether DDMC improves lesion detectability in PET-CT using FDG-18F.</p> Method <p>A moving platform simulated respiratory motion in a NEMA-IEC body phantom with varying amplitudes (0, 7, 10, 20, 30&#xa0;mm) and target-to-background ratios (2, 5, 10.5). Scans were reconstructed with and without DDMC, and the spherical targets’ maximal and mean recovery coefficient (RC) and contrast-to-noise ratio (CNR) were measured.</p> Results <p>DDMC results in higher RC values in the target spheres. CNR values increase for small, high-motion affected targets but decrease for larger spheres with smaller amplitudes. A sub-analysis shows that DDMC increases the contrast of the sphere along with a 36% increase in background noise.</p> Conclusion <p>While DDMC significantly enhances contrast (RC), its impact on detectability (CNR) is less profound due to increased background noise. CNR improves for small targets with high motion amplitude, potentially enhancing the detectability of low-uptake lesions. Given that the increased background noise may reduce detectability for targets unaffected by motion, we suggest that DDMC reconstructions are used best in addition to non-DDMC reconstructions.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Effect of data-driven motion correction for respiratory movement on lesion detectability in PET-CT: a phantom study

  • Marloes A. de Winter,
  • Robin Gevers,
  • Jules Lavalaye,
  • Jan B. A. Habraken,
  • Matteo Maspero

摘要

Purpose

While data-driven motion correction (DDMC) techniques have proven to enhance the visibility of lesions affected by motion, their impact on overall detectability remains unclear. This study investigates whether DDMC improves lesion detectability in PET-CT using FDG-18F.

Method

A moving platform simulated respiratory motion in a NEMA-IEC body phantom with varying amplitudes (0, 7, 10, 20, 30 mm) and target-to-background ratios (2, 5, 10.5). Scans were reconstructed with and without DDMC, and the spherical targets’ maximal and mean recovery coefficient (RC) and contrast-to-noise ratio (CNR) were measured.

Results

DDMC results in higher RC values in the target spheres. CNR values increase for small, high-motion affected targets but decrease for larger spheres with smaller amplitudes. A sub-analysis shows that DDMC increases the contrast of the sphere along with a 36% increase in background noise.

Conclusion

While DDMC significantly enhances contrast (RC), its impact on detectability (CNR) is less profound due to increased background noise. CNR improves for small targets with high motion amplitude, potentially enhancing the detectability of low-uptake lesions. Given that the increased background noise may reduce detectability for targets unaffected by motion, we suggest that DDMC reconstructions are used best in addition to non-DDMC reconstructions.