Background <p>The topic of the effect of the patient table on attenuation in myocardial perfusion imaging (MPI) SPECT is gaining new relevance due to deep learning methods. Existing studies on this effect are old, rare and only consider phantom measurements, not patient studies. This study investigates the effect of the patient table on attenuation based on the difference between reconstructions of phantom scans and polar maps of patient studies.</p> Methods <p>Jaszczak phantom scans are acquired according to quality control and MPI procedures. An algorithm is developed to automatically remove the patient table from the CT for attenuation correction. The scans are then reconstructed with attenuation correction either with or without the patient table in the CT. The reconstructions are compared qualitatively and on the basis of their percentage difference. In addition, a small retrospective cohort of 15 patients is examined by comparing the resulting polar maps. Polar maps are compared qualitatively and based on the segment perfusion scores.</p> Results <p>The phantom reconstructions look qualitatively similar in both the quality control and MPI procedures. The percentage difference is highest in the lower part of the phantom, but it always remains below 17.5%. Polar maps from patient studies also look qualitatively similar. Furthermore, the segment scores are not significantly different (<i>p</i>=0.83).</p> Conclusions <p>The effect of the patient table on attenuation in MPI SPECT is negligible.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

On the Effect of the Patient Table on Attenuation in Myocardial Perfusion Imaging SPECT

  • Tamino Huxohl,
  • Gopesh Patel,
  • Wolfgang Burchert

摘要

Background

The topic of the effect of the patient table on attenuation in myocardial perfusion imaging (MPI) SPECT is gaining new relevance due to deep learning methods. Existing studies on this effect are old, rare and only consider phantom measurements, not patient studies. This study investigates the effect of the patient table on attenuation based on the difference between reconstructions of phantom scans and polar maps of patient studies.

Methods

Jaszczak phantom scans are acquired according to quality control and MPI procedures. An algorithm is developed to automatically remove the patient table from the CT for attenuation correction. The scans are then reconstructed with attenuation correction either with or without the patient table in the CT. The reconstructions are compared qualitatively and on the basis of their percentage difference. In addition, a small retrospective cohort of 15 patients is examined by comparing the resulting polar maps. Polar maps are compared qualitatively and based on the segment perfusion scores.

Results

The phantom reconstructions look qualitatively similar in both the quality control and MPI procedures. The percentage difference is highest in the lower part of the phantom, but it always remains below 17.5%. Polar maps from patient studies also look qualitatively similar. Furthermore, the segment scores are not significantly different (p=0.83).

Conclusions

The effect of the patient table on attenuation in MPI SPECT is negligible.