Deep learning based bilateral filtering for edge-preserving denoising of respiratory-gated PET
摘要
Residual image noise is substantial in positron emission tomography (PET) and one of the factors limiting lesion detection, quantification, and overall image quality. Thus, improving noise reduction remains of considerable interest. This is especially true for respiratory-gated PET investigations. The only broadly used approach for noise reduction in PET imaging has been the application of low-pass filters, usually Gaussians, which however leads to loss of spatial resolution and increased partial volume effects affecting detectability of small lesions and quantitative data evaluation. The bilateral filter (BF) — a locally adaptive image filter — allows to reduce image noise while preserving well defined object edges but manual optimization of the filter parameters for a given PET scan can be tedious and time-consuming, hampering its clinical use. In this work we have investigated to what extent a suitable deep learning based approach can resolve this issue by training a suitable network with the target of reproducing the results of manually adjusted case-specific bilateral filtering.
MethodsAltogether, 69 respiratory-gated clinical PET/CT scans with three different tracers (
The automated procedure used for filter parameter determination chose adequate filter parameters for the majority of the data sets with only 19 patient data sets requiring manual tuning. Evaluation of the focal uptake ROIs revealed that CNN as well as BF based filtering essentially maintain the focal
Our results show that a neural network based denoising can reproduce the results of a case by case optimized BF in a fully automated way. Apart from rare cases it led to images of practically identical quality regarding noise level, edge preservation, and signal recovery. We believe such a network might proof especially useful in the context of improved motion correction of respiratory-gated PET studies but could also help to establish BF-equivalent edge-preserving CNN filtering in clinical PET since it obviates time consuming manual BF parameter tuning.