11C-CFT PET brain imaging in Parkinson’s disease using a total-body PET/CT scanner

  • Xiaolin Sun,
  • Xiaoyue Tan,
  • Qing Zhang,
  • Shanzhen He,
  • Siyun Wang,
  • Yongrong Zhou,
  • Qi Huang,
  • Lei Jiang

摘要

Purpose

This study aimed to evaluate the feasibility of 11C-CFT PET brain imaging in Parkinson’s Disease using a total-body PET/CT scanner and explore the optimal scan duration to guide the clinical practice.

Methods

Thirty-two patients with Parkinson’s disease (PD) performing 11C-CFT PET/CT brain imaging using a total-body PET/CT scanner were retrospectively enrolled. The PET data acquired over a period of 900 s were reconstructed into groups of different durations: 900-s, 720-s, 600-s, 480-s, 300-s, 180-s, 120-s, 60-s, and 30-s (G900 to G30). The subjective image quality analysis was performed using 5-point scales. Semi-quantitative measurements were analyzed by SUVmean and dopamine transporter (DAT) binding of key brain regions implicated in PD, including the caudate nucleus and putamen. The full-time images (G900) were served as reference.

Results

The overall G900, G720, and G600 image quality scores were 5.0 ± 0.0, 5.0 ± 0.0, and 4.9 ± 0.3 points, respectively, and there was no significant difference among these groups (P > 0.05). A significant decrease in these scores at durations shorter than 600 s was observed when compared to G900 images (P < 0.05). However, all G300 image quality was clinically acceptable (≥ 3 points). As the scan duration reduced, the SUVmean and DAT binding of caudate nucleus and putamen decreased progressively, while there were no statistically significant variations in the SUVmean of the background among the different groups. Moreover, the changes in the lesion DAT binding (ΔDAT-binding) between the full-time reference G900 image and other reconstructed group G720 to G30 images generally increased along with the reduced scan time.

Conclusion

Sufficient image quality and lesion conspicuity could be achieved at 600-s scan duration for 11C-CFT PET brain imaging in PD assessment using a total-body PET/CT scanner, while the image quality of G300 was acceptable to meet clinical diagnosis, contributing to improve patient compliance and throughput of PET brain imaging.