Evaluating a radiotherapy deep learning synthetic CT algorithm for PET-MR attenuation correction in the pelvis
摘要
Positron emission tomography–magnetic resonance (PET-MR) attenuation correction is challenging because the MR signal does not represent tissue density and conventional MR sequences cannot image bone. A novel zero echo time (ZTE) MR sequence has been previously developed which generates signal from cortical bone with images acquired in 65 s. This has been combined with a deep learning model to generate a synthetic computed tomography (sCT) for MR-only radiotherapy. This study aimed to evaluate this algorithm for PET-MR attenuation correction in the pelvis.
MethodsTen patients being treated with ano-rectal radiotherapy received a
Mean whole-image SUV differences were −0.02% (sCTAC) compared to −3.0% (MRAC), with larger differences in the bone regions (−0.5% to −16.3%). There was no difference in thresholded GTVs, with Dice similarity coefficients
Attenuation correction using this radiotherapy ZTE-based sCT algorithm was substantially more accurate than current MRAC methods with only a 40 s increase in MR acquisition time. This did not impact tumour delineation but did significantly improve the accuracy of whole-image and tumour SUV measurements, which were clinically equivalent to CTAC. This suggests PET images reconstructed with sCTAC would enable accurate quantitative PET images to be acquired on a PET-MR scanner.