Background <p>Biodegradable polymer (BP) and polymer-free (PF) drug eluting stents were designed to reduce late adverse events linked to durable-polymer coatings. However, concerns remain regarding the efficacy of PF–DES particularly due to variations in drug-release mechanisms.</p> Methods <p>RCTs comparing PF–DES and BP–DES in patients undergoing percutaneous coronary interventions were included. Primary endpoints were Target lesion revascularization (TLR), Target vessel revascularization (TVR), definite stent thrombosis, Cardiac death (CD), angiographic outcomes and optical coherence tomography (OCT) findings. Risk of bias was assessed using Cochrane’s ROB 2.0 and meta-analysis performed using RevMan 5.4.</p> Results <p>Ten RCTs including 9020 patients (4043 PF–DES; 4977 BP–DES) were included. At 12&#xa0;months, PF–DES showed a significantly higher TLR rate (2.08% vs. 1.36%; RR 1.55, 95% CI 1.087–2.21; <i>P</i> = 0.02), persisting at 24&#xa0;months (RR = 2.01, 95% CI 1.46–2.77, <i>P</i> &lt; 0.0001). At 1–3&#xa0;months, OCT-derived analyses demonstrated no statistically significant differences in uncovered struts or neointimal thickness between stent types, while findings related to malapposed struts were heterogenous and dependant on sensitivity analyses.</p> <p>Angiographically, PF–DES showed greater in-stent late lumen loss at 6–8&#xa0;months ([MD]: 0.24&#xa0;mm, 95% CI 0.17–0.30, <i>P</i> &lt; 0.00001), whereas minimal lumen diameter was comparable.</p> <p>However, no significant differences were found in target vessel revascularization (TVR), stent thrombosis, myocardial infarction, or mortality at 1- or 2-year follow-up.</p> Conclusions <p>PF–DES were associated with comparable safety to BP–DES; however, it exhibited higher rates of target lesion revascularization and greater late lumen loss were observed. OCT-derived findings were heterogenous and should be interpreted cautiously. Further RCTs with standardized imaging protocols and longer follow-up are warranted.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Safety, efficacy, OCT-assessed vascular healing and angiographic outcomes of polymer-free versus biodegradable-polymer drug-eluting stents in patients undergoing percutaneous coronary intervention: a systematic review and meta-analysis

  • Merna M. Abouelenien,
  • Mennatullah E. Mekky,
  • Zeyad Bakry,
  • Yazan Sawafta

摘要

Background

Biodegradable polymer (BP) and polymer-free (PF) drug eluting stents were designed to reduce late adverse events linked to durable-polymer coatings. However, concerns remain regarding the efficacy of PF–DES particularly due to variations in drug-release mechanisms.

Methods

RCTs comparing PF–DES and BP–DES in patients undergoing percutaneous coronary interventions were included. Primary endpoints were Target lesion revascularization (TLR), Target vessel revascularization (TVR), definite stent thrombosis, Cardiac death (CD), angiographic outcomes and optical coherence tomography (OCT) findings. Risk of bias was assessed using Cochrane’s ROB 2.0 and meta-analysis performed using RevMan 5.4.

Results

Ten RCTs including 9020 patients (4043 PF–DES; 4977 BP–DES) were included. At 12 months, PF–DES showed a significantly higher TLR rate (2.08% vs. 1.36%; RR 1.55, 95% CI 1.087–2.21; P = 0.02), persisting at 24 months (RR = 2.01, 95% CI 1.46–2.77, P < 0.0001). At 1–3 months, OCT-derived analyses demonstrated no statistically significant differences in uncovered struts or neointimal thickness between stent types, while findings related to malapposed struts were heterogenous and dependant on sensitivity analyses.

Angiographically, PF–DES showed greater in-stent late lumen loss at 6–8 months ([MD]: 0.24 mm, 95% CI 0.17–0.30, P < 0.00001), whereas minimal lumen diameter was comparable.

However, no significant differences were found in target vessel revascularization (TVR), stent thrombosis, myocardial infarction, or mortality at 1- or 2-year follow-up.

Conclusions

PF–DES were associated with comparable safety to BP–DES; however, it exhibited higher rates of target lesion revascularization and greater late lumen loss were observed. OCT-derived findings were heterogenous and should be interpreted cautiously. Further RCTs with standardized imaging protocols and longer follow-up are warranted.