Inter-lung asymmetrical airway closure cause insufflation delay between lungs in acute hypoxemic respiratory failure
摘要
Electrical Impedance Tomography (EIT) can quantify ventilation in the two lungs and be used to measure the airway opening pressure (AOP) of each lung. Asymmetrical AOPs can cause inter-lung insufflation delay.
ObjectivesTo assess the relation between AOP asymmetry and inter-lung insufflation delay at different PEEP levels.
MethodsPatients with acute hypoxemic respiratory failure and airway closure were included. Low-flow pressure-volume curves and EIT signal were recorded during controlled ventilation and for some patients in pressure support ventilation.
Results23 patients were studied, 22 patients had ARDS, 9 patients had asymmetrical airway closure with an AOP of 10 [
In asymmetrical airway closure, titrating PEEP can minimize inter-lung insufflation delay and can be monitored by EIT. Reducing the delay and reducing ventilation asymmetry is also feasible during pressure support ventilation when low flow inflation curves cannot be performed.