Background <p>Biocontainment protocols are critical for conducting infectious disease (ID) research, particularly when using small animal models in biosafety level (BSL) 2/3 environments. This study evaluates the impact of poly-methyl methacrylate (PMMA) containment vessels on the performance of preclinical positron emission tomography (PET) systems. We tested containment vessels designed with varying wall thicknesses (3, 6, and 9&#xa0;mm) to simulate ID imaging facility equipment and protocols. Through the use of multicomponent phantoms and in vivo mouse models of <i>Staphylococcus aureus</i> infection, we assessed key performance metrics including count rate, image quality, activity recovery, and spatial resolution.</p> Results <p>The results indicate that the use of PMMA containment causes only minor reductions in imaging performance. The thickest PMMA (9&#xa0;mm) led to a maximum 6.8% decrease in count rate, which remains well within the acceptable range of variation. Effects on spatial resolution were most noticeable for smaller structures within the phantom study, with a 19.65% difference in full width at half maximum (FWHM) for the thickest walled vessel. In vivo, using infected mice, the containment devices had modest effects on the task of activity concentration to be detected at the infection site, even with the thickest PMMA tube.</p> Conclusion <p>These findings suggest that PMMA biocontainment vessels have small but measurable impact on preclinical PET system performance, making them a viable and cost-effective solution for conducting infectious disease imaging under BSL-2/3 conditions. Specifically, the thinnest containment (3&#xa0;mm) had only minor effects on all tested parameters, suggesting it is well-suited for use in ID enclosures while maintaining accurate qualitative and quantitative assessments. This approach may reduce the burden for fully separate and specialized modifications for BSL-3 imaging facilities, and can be broadly applied to preclinical research involving pathogenic organisms.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Impact of biocontainment on small animal PET performance adapted for BSL-2/3 infectious disease imaging research

  • Isabella Salerno,
  • Nadia Benabdallah,
  • Amanda Fears,
  • Ryan Unnerstall,
  • Lindsey Hauck,
  • Sergey Komarov,
  • Linda Cox,
  • Hanwen Zhang,
  • Kevin Poenicke,
  • Joseph Aromando III,
  • Yuan-Chun Tai,
  • Timothy Wencewicz,
  • Deborah J. Veis,
  • Daniel L. J. Thorek

摘要

Background

Biocontainment protocols are critical for conducting infectious disease (ID) research, particularly when using small animal models in biosafety level (BSL) 2/3 environments. This study evaluates the impact of poly-methyl methacrylate (PMMA) containment vessels on the performance of preclinical positron emission tomography (PET) systems. We tested containment vessels designed with varying wall thicknesses (3, 6, and 9 mm) to simulate ID imaging facility equipment and protocols. Through the use of multicomponent phantoms and in vivo mouse models of Staphylococcus aureus infection, we assessed key performance metrics including count rate, image quality, activity recovery, and spatial resolution.

Results

The results indicate that the use of PMMA containment causes only minor reductions in imaging performance. The thickest PMMA (9 mm) led to a maximum 6.8% decrease in count rate, which remains well within the acceptable range of variation. Effects on spatial resolution were most noticeable for smaller structures within the phantom study, with a 19.65% difference in full width at half maximum (FWHM) for the thickest walled vessel. In vivo, using infected mice, the containment devices had modest effects on the task of activity concentration to be detected at the infection site, even with the thickest PMMA tube.

Conclusion

These findings suggest that PMMA biocontainment vessels have small but measurable impact on preclinical PET system performance, making them a viable and cost-effective solution for conducting infectious disease imaging under BSL-2/3 conditions. Specifically, the thinnest containment (3 mm) had only minor effects on all tested parameters, suggesting it is well-suited for use in ID enclosures while maintaining accurate qualitative and quantitative assessments. This approach may reduce the burden for fully separate and specialized modifications for BSL-3 imaging facilities, and can be broadly applied to preclinical research involving pathogenic organisms.