Optimizing ultrafast dynamic contrast-enhanced MRI scan duration in the differentiation of benign and malignant breast lesions
摘要
To determine the optimal scan duration for ultrafast DCE-MRI in effectively differentiating benign from malignant breast lesions.
MethodsThe study prospectively recruited participants who underwent breast ultrafast DCE-MRI from September 2021 to March 2023. A 30-phase breast ultrafast DCE-MRI on a 3.0-T MRI system was conducted with a 4.5-s temporal resolution. Scan durations ranged from 40.5 s to 135.0 s, during which the analysis is performed at three-phase intervals, forming eight dynamic sets (scan duration [SD]40.5s: 40.5 s, SD54s: 54.0 s, SD67.5s: 67.5 s, SD81s: 81.0 s, SD94.5s: 94.5 s, SD108s: 108.0 s, SD121.5s: 121.5 s, and SD135s: 135.0 s). Two ultrafast DCE-MRI parameters, maximum slope (MS) and initial area under the curve in 60 s (iAUC), were calculated for each dynamic set and compared between benign and malignant lesions. Areas under the receiver operating characteristic curve (AUCs) were used to assess their diagnostic performance.
ResultsA total of 140 women (mean age, 47 ± 11 years) with 151 lesions were included. MS and iAUC from eight dynamic sets exhibited significant differences between benign and malignant lesions (all p < 0.05), except iAUC at SD40.5s. The AUC of MS (AUC = 0.804) and iAUC (AUC = 0.659) at SD67.5s were significantly higher than their values at SD40.5s (AUC = 0.606 and 0.516; corrected p < 0.05). No significant differences in AUCs for MS and iAUC were observed from SD67.5s to SD135s (all corrected p > 0.05).
ConclusionsUltrafast DCE-MRI with a 67.5-s scan duration appears optimal for effectively differentiating malignant from benign breast lesions.
Critical relevance statementBy evaluating scan durations (40.5–135 s) and analyzing two ultrafast DCE-MRI parameters, we found a scan duration of 67.5 s optimal for discriminating between these lesions and offering a balance between acquisition time and diagnostic efficacy.
Key PointsUltrafast DCE-MRI can effectively differentiate malignant from benign breast lesions. A minimum of 67.5-sec ultrafast DCE-MRI scan duration is required to differentiate benign and malignant lesions. Extending the scan duration beyond 67.5 s did not significantly improve diagnostic accuracy.