Effects of different VV ECMO blood flow rates on lung perfusion assessment by hypertonic saline bolus-based electrical impedance tomography

  • Hongling Zhang,
  • Yongran Wu,
  • Xuehui Gao,
  • Chengchao Peng,
  • Ruirui Li,
  • Azhen Wang,
  • Jiancheng Zhang,
  • Shiying Yuan,
  • Le Yang,
  • Xiaojing Zou,
  • You Shang

摘要

Objective

Our study aimed to investigate the effects of different extracorporeal membrane oxygenation (ECMO) blood flow rates on lung perfusion assessment using the saline bolus-based electrical impedance tomography (EIT) technique in patients on veno-venous (VV) ECMO.

Methods

In this single-centered prospective physiological study, patients on VV ECMO who met the ECMO weaning criteria were assessed for lung perfusion using saline bolus-based EIT at various ECMO blood flow rates (gradually decreased from 4.5 L/min to 3.5 L/min, 2.5 L/min, 1.5 L/min, and finally to 0 L/min). Lung perfusion distribution, dead space, shunt, ventilation/perfusion matching, and recirculation fraction at different flow rates were compared.

Results

Fifteen patients were included. As the ECMO blood flow rate decreased from 4.5 L/min to 0 L/min, the recirculation fraction decreased significantly. The main EIT-based findings were as follows. (1) Median lung perfusion significantly increased in region-of-interest (ROI) 2 and the ventral region [38.21 (34.93–42.16)% to 41.29 (35.32–43.75)%, p = 0.003, and 48.86 (45.53–58.96)% to 54.12 (45.07–61.16)%, p = 0.037, respectively], whereas it significantly decreased in ROI 4 and the dorsal region [7.87 (5.42–9.78)% to 6.08 (5.27–9.34)%, p = 0.049, and 51.14 (41.04–54.47)% to 45.88 (38.84–54.93)%, p = 0.037, respectively]. (2) Dead space significantly decreased, and ventilation/perfusion matching significantly increased in both the ventral and global regions. (3) No significant variations were observed in regional and global shunt.

Conclusions

During VV ECMO, the ECMO blood flow rate, closely linked to recirculation fraction, could affect the accuracy of lung perfusion assessment using hypertonic saline bolus-based EIT.