Background

New prosthetic valves and surgical approaches that shorten operation time and improve the outcome of patients with aortic valve (AV) infective endocarditis (IE) and AV insufficiency (AVI) are crucial. The aim of this study was to evaluate the outcome of patients with AV IE or AVI treated with the EDWARDS INTUITY Rapid-Deployment AV prosthesis for this off-label indication.

Methods

This single-centre retrospective study analyzed data from patients who underwent AV replacement with the EDWARDS INTUITY Rapid-Deployment AV prosthesis for AV IE or regurgitation. (n = 8 for IE and n = 6 for AVI).

Results

Heart-lung machine times were significantly shorter in the AVI group (111.3 ± 20.7 min) compared to the IE group (171.9 ± 52.4 min) (p = 0.02). Aortic cross-clamp followed a similar trend (73.7 ± 9.9 min for AVI vs. 113.4 ± 35.6 min for IE) (p = 0.02). The length of ICU stay was also shorter in the AVI group (3.8 ± 2.6 days) compared to the IE group (16.9 ± 8.9 days) (p = 0.005). Postoperative echocardiography revealed no paravalvular leakage or significant valvular dysfunction in any patient. One patient died postoperatively from aspiration pneumonia.

Conclusion

The INTUITY valve demonstrates as a safe option for complex AV IE and AVI surgery. Further prospective studies with larger patient cohorts are necessary to confirm these findings and explore the long-term benefits of this approach.