Background <p>This study reports the single-institution clinical experience of multifield pencil beam scanning (PBS) intensity-modulated proton therapy (IMPT) and dosimetric comparison to proton arc for uveal melanoma (UM) in a regular PBS gantry room.</p> Methods <p>Eleven consecutive UM patients were treated with IMPT to 50&#xa0;Gy in 5 fractions. A customized gaze-fixation device attached to the thermoplastic mask was used to reproduce the globe position for each patient. IMPT plans were robustly optimized with perturbations of 3&#xa0;mm setup and 3.5% range uncertainties using 3–4 fields without apertures. Each plan was robustly reoptimized (using the same perturbation parameters) using two non-coplanar arc fields in the RayStation treatment planning system. Treatment quality for both plans was evaluated daily using CBCT-generated synthetic CT. Target coverage, conformity, and mean/maximum doses to adjacent organs were assessed.</p> Results <p>Proton arc plans provided comparable plan quality compared to IMPT plans. Similar target coverage was achieved, with an average GTV D95% equal to 101.1% <InlineEquation ID="IEq1"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq1.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 1.0% and 101.4% <InlineEquation ID="IEq2"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq2.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 0.4% for IMPT and proton arc plans, respectively. Proton arc improves the conformity index (RTOG) compared to IMPT plans (average 0.96 <InlineEquation ID="IEq3"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq3.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 0.23 vs. 0.88 <InlineEquation ID="IEq4"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq4.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 0.18, <i>p</i> = 0.11). Both modalities met all the clinical goals for organs-at-risk (OARs), while proton arc significantly reduced the maximum dose for the retina from, on average, 54.5 <InlineEquation ID="IEq5"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq5.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 0.7 to 53.2 <InlineEquation ID="IEq6"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="13014_2025_2621_Article_IEq6.gif" Format="GIF" Height="13" Rendition="HTML" Resolution="72" Type="Linedraw" Width="18" /> </InlineMediaObject> <EquationSource Format="TEX">\(\:\pm\:\)</EquationSource> </InlineEquation> 0.3&#xa0;Gy (<i>p</i> &lt; 0.01). Treatment evaluation on synthetic CT showed that the doses received by patients were highly consistent with the planned doses, with a relative target coverage (D95%) difference within 3.5% for IMPT and 3.1% for proton arc, and the D95% of actual delivery exceeding 98.7% and 98.2%, respectively. The doses delivered to OARs did not exceed clinical constraints.</p> Conclusions <p>This is a novel report on proton arc for ocular tumors and gantry-based multifield PBS proton treatment for these tumors. This study demonstrated that both modalities can meet the clinical goals. The IMPT is currently clinically implanted, and 2-field non-coplanar proton arc plans can achieve comparable dosimetric metrics to those of IMPT plans when the deliver technique is matured.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Gantry-based pencil beam scanning proton therapy for uveal melanoma: IMPT versus proton arc therapy

  • Hang Qi,
  • Lei Hu,
  • Sheng Huang,
  • Yen-Po Lee,
  • Francis Yu,
  • Qing Chen,
  • Yunjie Yang,
  • Minglei Kang,
  • Huifang Zhai,
  • Milo Vermeulen,
  • Andy Shim,
  • Peter Park,
  • Xuanfeng Ding,
  • Jun Zhou,
  • David H. Abramson,
  • Jasmine H. Francis,
  • Charles B. Simone 2nd,
  • Christopher A. Barker,
  • Haibo Lin

摘要

Background

This study reports the single-institution clinical experience of multifield pencil beam scanning (PBS) intensity-modulated proton therapy (IMPT) and dosimetric comparison to proton arc for uveal melanoma (UM) in a regular PBS gantry room.

Methods

Eleven consecutive UM patients were treated with IMPT to 50 Gy in 5 fractions. A customized gaze-fixation device attached to the thermoplastic mask was used to reproduce the globe position for each patient. IMPT plans were robustly optimized with perturbations of 3 mm setup and 3.5% range uncertainties using 3–4 fields without apertures. Each plan was robustly reoptimized (using the same perturbation parameters) using two non-coplanar arc fields in the RayStation treatment planning system. Treatment quality for both plans was evaluated daily using CBCT-generated synthetic CT. Target coverage, conformity, and mean/maximum doses to adjacent organs were assessed.

Results

Proton arc plans provided comparable plan quality compared to IMPT plans. Similar target coverage was achieved, with an average GTV D95% equal to 101.1% \(\:\pm\:\) 1.0% and 101.4% \(\:\pm\:\) 0.4% for IMPT and proton arc plans, respectively. Proton arc improves the conformity index (RTOG) compared to IMPT plans (average 0.96 \(\:\pm\:\) 0.23 vs. 0.88 \(\:\pm\:\) 0.18, p = 0.11). Both modalities met all the clinical goals for organs-at-risk (OARs), while proton arc significantly reduced the maximum dose for the retina from, on average, 54.5 \(\:\pm\:\) 0.7 to 53.2 \(\:\pm\:\) 0.3 Gy (p < 0.01). Treatment evaluation on synthetic CT showed that the doses received by patients were highly consistent with the planned doses, with a relative target coverage (D95%) difference within 3.5% for IMPT and 3.1% for proton arc, and the D95% of actual delivery exceeding 98.7% and 98.2%, respectively. The doses delivered to OARs did not exceed clinical constraints.

Conclusions

This is a novel report on proton arc for ocular tumors and gantry-based multifield PBS proton treatment for these tumors. This study demonstrated that both modalities can meet the clinical goals. The IMPT is currently clinically implanted, and 2-field non-coplanar proton arc plans can achieve comparable dosimetric metrics to those of IMPT plans when the deliver technique is matured.