Preliminary study of feasibility of surface-guided radiotherapy with concurrent tumor treating fields for glioblastoma: region of interest
摘要
To evaluate the impact of the residual setup errors from differently shaped region of interest (ROI) and investigate if surface-guided setup can be used in radiotherapy with concurrent tumor treating fields (TTFields) for glioblastoma.
MethodsFifteen patients undergone glioblastoma radiotherapy with concurrent TTFields were involved. Firstly, four shapes of region of interest (ROI) (strip-shaped, T-shaped,
The algebraic sum of the average residual setup errors in six degrees (mm in translational directions and ° in rotational directions) of the four types were 6.9, 1.1, 4.1 and 3.5 respectively. In terms of the ROI-shape-dependent dosimetry deviations, the D98% of PTV dropped off by (3.4 ± 2.0)% (p < 0.05), (0.3 ± 0.5)% (p < 0.05), (0.9 ± 0.9)% (p < 0.05) and (1.1 ± 0.8)% (p < 0.05). The D98% of CTV dropped off by (0.5 ± 0.6)% (p < 0.05) for the strip-shaped ROI while remained unchanged for others.
ConclusionSurface-guided setup is feasible in radiotherapy with concurrent TTFields and a medium-sized T-shaped ROI is appropriate for the surface-based guidance.