Effect of frozen storage conditions on antithrombin protein C and protein S activity assay stability

  • Houmei Feng,
  • Danyu Song,
  • Qiang Xu,
  • Xiaohui Cai,
  • Jianru Liu,
  • Yang Zhang,
  • Zhou Zhou

摘要

Background

Inherited antithrombin, protein C, and protein S deficiency increase the risk of venous thromboembolism. The presence of defects can be identified by clinical laboratory assays. In most Chinese clinical laboratories, the screening tests for antithrombin, protein C, and protein S deficiency are their activity assays. Ensuring appropriate pre-analytical storage conditions for activity tests is essential. This study aimed to assess the effects of storage conditions on antithrombin, protein C, and protein S activity in frozen plasma.

Methods

We collected the remaining plasma of 29 patients. The baseline of antithrombin, protein C, and protein S activity values were tested within 4 h. Then, each sample was sub-packaged into 4 EP tubes, and was stored at -20 °C for 3 days, -20 °C for 7 days, -80 °C for 3 days, and − 80 °C for 7 days, respectively. After thawing, samples were tested by two systems.

Results

The percentage deviation of antithrombin and protein C activity assay was<10% compared with the initial values. Protein S activity showed a significant reduction in frozen plasma, with a deviation > 10%. Some samples, initially within the normal range, were classified as abnormal after freezing storage.

Conclusions

Our study indicated that antithrombin and protein C remain stable when stored at -20 °C or -80 °C in a week. We argued that Protein S activity is not stable in frozen plasma. The use of frozen-thawed plasma for PS activity assay may result in overdiagnosis of protein S deficiency.