错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Rapid detection of ceftriaxone-resistant Salmonella by matrix-assisted laser desorption–ionization time-of-flight mass spectrometry combined with the ratio of optical density

  • Chao Fang,
  • Zheng Zhou,
  • Mingming Zhou,
  • Jianping Li

摘要

Background

The increased resistance rate of Salmonella to third-generation cephalosporins represented by ceftriaxone (CRO) may result in the failure of the empirical use of third-generation cephalosporins for the treatment of Salmonella infection in children. The present study was conducted to evaluate a novel method for the rapid detection of CRO-resistant Salmonella (CRS).

Methods

We introduced the concept of the ratio of optical density (ROD) with and without CRO and combined it with matrix-assisted laser desorption–ionization time-of-flight mass spectrometry (MALDI-TOF MS) to establish a new protocol for the rapid detection of CRS.

Results

The optimal incubation time and CRO concentration determined by the model strain test were 2 h and 8 µg/ml, respectively. We then conducted confirmatory tests on 120 clinical strains. According to the receiver operating characteristic curve analysis, the ROD cutoff value for distinguishing CRS and non-CRS strains was 0.818 [area under the curve: 1.000; 95% confidence interval: 0.970–1.000; sensitivity: 100.00%; specificity: 100%; P < 10− 3].

Conclusions

In conclusion, the protocol for the combined ROD and MALDI-TOF MS represents a rapid, accurate, and economical method for the detection of CRS.