Evaluating the relationship between clinician experience and accuracy in robotic-assisted dental implant placement: a retrospective study
摘要
This study aims to evaluate the accuracy of Level of Autonomy 2 (LOA2) robotic-assisted implant placement by dentists with varying experience, analyze learning curves, and identify accuracy-related risk factors.
Materials and methodsThis retrospective study included 362 patients (585 implants) who underwent LOA2 robotic-assisted implant placement (April 2022–April 2024). Six novice clinicians (no robotic experience) and two expert clinicians (> 50 robotic surgeries) were included. Seven deviation parameters were assessed by comparing planned and actual implant positions. Thirteen potential risk factors were evaluated using statistical analyses in R.
ResultsAfter extraoral robotic training, novice clinicians achieved accuracy comparable to experts, with minor exceptions: vertical platform deviation (0.373 ± 0.566 mm vs. 0.255 ± 0.438 mm, p = 0.007) and vertical apex deviation (0.348 ± 0.488 mm vs. 0.249 ± 0.437 mm, p = 0.009). Learning curves for both groups remained flat (coefficient < 0.001), with no significant accuracy improvement over cases increase. Key risk factors for increased deviations included: immediate/early implantation; implant length > 10 mm; poor bone quality; maxillary placement (p < 0.05), and machined-related factors.
ConclusionsThe LOA 2 robotic-assisted implant system minimizes the dependence of implant placement accuracy on clinician experience, with only minor differences in vertical deviations between novices and experts. Even though accuracy remains at a high level, high deviations are associated not only with traditional risk factors but also with machine-related risks.