Background <p>Traditional and innovative materials are widely used in dentistry; however, the mechanical behavior of hybrid custom implant abutments, particularly stress distribution in various material combinations, is not fully understood. The study aims to evaluate the mechanical behavior of hybrid custom implant abutments made from various material combinations, including their effects on von Mises stress, maximum and minimum principal stresses, and deformation.</p> Methods <p>Two 3-dimensional (3D) models were constructed: a 1.5&#xa0;mm subcrestal as the test and an equicrestal model as the control. The subcrestal model explored seven materials (Zirconia, Titanium, Lithium Disilicate, Polymer-Infiltrated Ceramic Networks, PEEK, PEEK reinforced with carbon fiber and reinforced with glass fiber) in various abutment and crown combinations. Each model included an implant, titanium base abutment, abutment screw, a custom abutment, a zirconia crown, and bone. A 200&#xa0;N load was applied, and a Finite Element Analysis (FEA) assessed peak, volume average, and distribution of von Mises stress and principal stress.</p> Results <p>The titanium base (Tibase) exhibited the highest peak and volume average von Mises stresses (306–429&#xa0;MPa), followed by the custom abutment (40–95&#xa0;MPa) and crown (46–81&#xa0;MPa). Material changes significantly impacted stress distribution in the Tibase and customized abutments. PICN, Zirconia, Titanium, and Lithium Disilicate abutments showed peak principal stresses between 77 and 85&#xa0;MPa, while PEEK variants reduced stress in the custom abutment (35–66&#xa0;MPa) but increased it in the Ti-base (356–405&#xa0;MPa). PEEK also increased minimum principal stresses in the Ti-base (-400 to -600&#xa0;MPa).</p> Conclusions <p>Abutment materials have a greater impact on stress outcomes compared to crown materials. Abutments with high Young’s modulus contribute to increased core system stiffness in hybrid custom abutment complexes. Choosing abutment materials with a high Young’s modulus for hybrid custom implant abutments is essential to optimize stress distribution and enhance the stability of the implant system.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Mechanical behavior of hybrid custom implant abutments with various crown materials: a 3D finite element analysis

  • Pongsakorn Poovarodom,
  • Guilherme Faria Moura,
  • Fabio Antonio Piola Rizzante,
  • Chaiy Rungsiyakull,
  • Jarupol Suriyawanakul,
  • Pimduen Rungsiyakull

摘要

Background

Traditional and innovative materials are widely used in dentistry; however, the mechanical behavior of hybrid custom implant abutments, particularly stress distribution in various material combinations, is not fully understood. The study aims to evaluate the mechanical behavior of hybrid custom implant abutments made from various material combinations, including their effects on von Mises stress, maximum and minimum principal stresses, and deformation.

Methods

Two 3-dimensional (3D) models were constructed: a 1.5 mm subcrestal as the test and an equicrestal model as the control. The subcrestal model explored seven materials (Zirconia, Titanium, Lithium Disilicate, Polymer-Infiltrated Ceramic Networks, PEEK, PEEK reinforced with carbon fiber and reinforced with glass fiber) in various abutment and crown combinations. Each model included an implant, titanium base abutment, abutment screw, a custom abutment, a zirconia crown, and bone. A 200 N load was applied, and a Finite Element Analysis (FEA) assessed peak, volume average, and distribution of von Mises stress and principal stress.

Results

The titanium base (Tibase) exhibited the highest peak and volume average von Mises stresses (306–429 MPa), followed by the custom abutment (40–95 MPa) and crown (46–81 MPa). Material changes significantly impacted stress distribution in the Tibase and customized abutments. PICN, Zirconia, Titanium, and Lithium Disilicate abutments showed peak principal stresses between 77 and 85 MPa, while PEEK variants reduced stress in the custom abutment (35–66 MPa) but increased it in the Ti-base (356–405 MPa). PEEK also increased minimum principal stresses in the Ti-base (-400 to -600 MPa).

Conclusions

Abutment materials have a greater impact on stress outcomes compared to crown materials. Abutments with high Young’s modulus contribute to increased core system stiffness in hybrid custom abutment complexes. Choosing abutment materials with a high Young’s modulus for hybrid custom implant abutments is essential to optimize stress distribution and enhance the stability of the implant system.