Removal of broken abutment screws using ultrasonic tip – a heat development in-vitro study

  • Vasilios Alevizakos,
  • Anna-Lena Bergmann,
  • Constantin von See

摘要

Background

Dental implants can cause complications, including the loosening of the abutment screw or fracture. However, there is no standardized technique for removing broken abutment screws. This necessitates further research.

Objective

This study aimed to measure heat generation during screw removal to better understand its implications for dental implant procedures.

Material and methods

The experimental setup involved using synthetic bone blocks and titanium implants. An ultrasonically operated instrument tip was utilized for screw removal. Infrared thermometry was employed for accurate temperature measurement, considering factors such as emissivity and distance. Statistical analysis using linear regression and ANOVA was conducted.

Results

The findings revealed an initial rapid temperature increase during the removal process, followed by a gradual decrease. The regression model demonstrated a strong correlation between time and temperature, indicating the heat generation pattern.

Conclusion

Heat generation during screw removal poses risks such as tissue damage and integration issues. Clinicians should minimize heat risks through an intermittent approach. The lack of a standardized technique requires further research and caution. Understanding the generated heat optimizes implant procedures.