Background <p>Pedicle screws are commonly used in spinal surgeries, but screw loosening remains a major concern, even in non-osteoporotic patients. Predicting pedicle screw stability via the insertion torque is a controversial topic, mainly studied on osteoporotic cadavers. Whether the insertion torque is suitable for patients with healthy bone mineral density (BMD) remains unknown. The aim was to investigate the influencing factors, namely insertion torque, BMD, screw diameter, length, surface area, volume, screw-in rotations, vertebral level, on the screw loosening stability during distractions and to understand if intra-operative predictions are possible.</p> Methods <p>Non-osteoporotic thoraco-lumbar vertebrae (<i>n</i> = 50) were used to implant five different pedicle screws (<i>n</i> = 100) while measuring the insertion torque. After embedding the endplates, the force needed to distract the screw head by 1&#xa0;mm was tested.</p> Results <p>The insertion toque (2.3 ± 0.9 Nm) showed the highest influence on the distraction force (324.8 ± 84.4&#xa0;N) followed by the screw size and vertebral level. BMD did not show any effects.</p> Conclusions <p>The linear correlation of insertion torque and the bending force suggests an alternative prediction metric for screw loosening which could improve the outcome of surgeries and patients’ safety. This is potentially a simple, intra-operative method, which can be used in future.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Prediction of screw loosening by measuring the insertion torque in non-osteoporotic patients: an in vitro study

  • Jan Ulrich Jansen,
  • Laura Zengerle,
  • Carsten Hackenbroch,
  • Jens Dreyhaupt,
  • Youping Tao,
  • Hans-Joachim Wilke

摘要

Background

Pedicle screws are commonly used in spinal surgeries, but screw loosening remains a major concern, even in non-osteoporotic patients. Predicting pedicle screw stability via the insertion torque is a controversial topic, mainly studied on osteoporotic cadavers. Whether the insertion torque is suitable for patients with healthy bone mineral density (BMD) remains unknown. The aim was to investigate the influencing factors, namely insertion torque, BMD, screw diameter, length, surface area, volume, screw-in rotations, vertebral level, on the screw loosening stability during distractions and to understand if intra-operative predictions are possible.

Methods

Non-osteoporotic thoraco-lumbar vertebrae (n = 50) were used to implant five different pedicle screws (n = 100) while measuring the insertion torque. After embedding the endplates, the force needed to distract the screw head by 1 mm was tested.

Results

The insertion toque (2.3 ± 0.9 Nm) showed the highest influence on the distraction force (324.8 ± 84.4 N) followed by the screw size and vertebral level. BMD did not show any effects.

Conclusions

The linear correlation of insertion torque and the bending force suggests an alternative prediction metric for screw loosening which could improve the outcome of surgeries and patients’ safety. This is potentially a simple, intra-operative method, which can be used in future.