Background <p>This study aimed to evaluate the accuracy and reproducibility of a newly developed portable and radiation-free three-dimensional spine sensing system (3D-SSS) for scoliosis assessment.</p> Methods <p>A total of 145 patients underwent full-spine imaging using the EOS imaging system, and 3D-SSS data were collected between February 2023 and April 2023. A radiologist used sterEOS software to reconstruct the spine in 3D and obtain the Cobb angle. One radiologist and one orthopedist independently measured the patients using 3D-SSS, with the orthopedist performing two measurements per patient. The 3D-SSS post-processing system automatically generated the Cobb angle.</p> Results <p>The mean Cobb angles obtained from EOS and 3D-SSS were 13.7 ± 9.9° (0.5∽45.7°) and 12.5 ± 8.6° (0.4∽40°), respectively. The intraclass correlation coefficient (ICC) for reliability between EOS and 3D-SSS was 0.921, indicating excellent agreement. Bland–Altman analysis revealed a bias of -1.171° between EOS and 3D-SSS, with only 10 patients outside the limits of agreement (-8.3∽6.0°). The root mean square error between EOS and 3D-SSS was 3.2°. A strong correlation was observed between the Cobb angles measured by EOS and 3D-SSS (<i>r</i> = 0.931, <i>P</i> &lt; 0.001). The receiver operating characteristics curve showed that the diagnostic performance of 3D-SSS for scoliosis was 0.953 (<i>P</i> &lt; 0.001). The sensitivity, specificity, positive predictive value, and negative predictive value of 3D-SSS for diagnosing scoliosis were 87.8%, 92.1%, 93.5%, and 85.3%, respectively. The intraobserver and interobserver ICCs for Cobb angles derived from 3D-SSS were 0.969 and 0.934, respectively, demonstrating excellent reproducibility.</p> Conclusions <p>The portable and radiation-free 3D-SSS accurately measured scoliosis and provided highly reproducible data. This system offers a novel method for clinicians to screen and monitor scoliosis in young patients.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

A novel portable and radiation-free method for assessing scoliosis: an accurate and reproducible study

  • Hui Wang,
  • Yunfeng Zhu,
  • Qiyuan Bao,
  • Yong Lu,
  • Fuhua Yan,
  • Lianjun Du,
  • Le Qin

摘要

Background

This study aimed to evaluate the accuracy and reproducibility of a newly developed portable and radiation-free three-dimensional spine sensing system (3D-SSS) for scoliosis assessment.

Methods

A total of 145 patients underwent full-spine imaging using the EOS imaging system, and 3D-SSS data were collected between February 2023 and April 2023. A radiologist used sterEOS software to reconstruct the spine in 3D and obtain the Cobb angle. One radiologist and one orthopedist independently measured the patients using 3D-SSS, with the orthopedist performing two measurements per patient. The 3D-SSS post-processing system automatically generated the Cobb angle.

Results

The mean Cobb angles obtained from EOS and 3D-SSS were 13.7 ± 9.9° (0.5∽45.7°) and 12.5 ± 8.6° (0.4∽40°), respectively. The intraclass correlation coefficient (ICC) for reliability between EOS and 3D-SSS was 0.921, indicating excellent agreement. Bland–Altman analysis revealed a bias of -1.171° between EOS and 3D-SSS, with only 10 patients outside the limits of agreement (-8.3∽6.0°). The root mean square error between EOS and 3D-SSS was 3.2°. A strong correlation was observed between the Cobb angles measured by EOS and 3D-SSS (r = 0.931, P < 0.001). The receiver operating characteristics curve showed that the diagnostic performance of 3D-SSS for scoliosis was 0.953 (P < 0.001). The sensitivity, specificity, positive predictive value, and negative predictive value of 3D-SSS for diagnosing scoliosis were 87.8%, 92.1%, 93.5%, and 85.3%, respectively. The intraobserver and interobserver ICCs for Cobb angles derived from 3D-SSS were 0.969 and 0.934, respectively, demonstrating excellent reproducibility.

Conclusions

The portable and radiation-free 3D-SSS accurately measured scoliosis and provided highly reproducible data. This system offers a novel method for clinicians to screen and monitor scoliosis in young patients.