A M-PEEK rod system to stabilize spinal motion after graded facetectomy: a finite element study

  • Yi-An Li,
  • Shih-Liang Shih,
  • Hsin-Chang Chen

摘要

Background

Resecting the facet joint to relieve nerve pain can lead to spinal instability, deformity, and abnormal pressure on the anterior of the intravertebral disc. To mitigate these issues, surgeons often limit the amount of bone removed during facetectomy or stabilize the spine by fusion to maintain lumbar stability. This study aimed to assess how a M-PEEK rod system influenced the stability of the lumbar spine during graded facetectomy.

Methods

Facetectomy was performed on a validated L1-L5 finite element model which was then simulated both with and without the M-PEEK rod system.

Results

In extension, models implanted with M-PEEK in the interspinous space of L3/L4 experienced a 35.2% decrease in range of motion (ROM) at L3/L4, while others saw an 8.4–24.8% increase. For axial rotation, the ROM at L3/L4 increased by 2.2–5.4% in models with the M-Rod, and by 4.9–12.9% in models without the implant. In lateral flexion, the ROM at L3/L4 increased by 8.4–14.3% in models without a PEEK M-Rod (facetectomy only), with adjacent segments experiencing a 6.5% decrease in ROM in the implanted models. Overall, the difference in ROM between the intact and implanted models was minimal.

Conclusions

Facetectomy involving the removal of 50% or more of the facet joint significantly increases range of motion and maximum intradiscal pressure, potentially accelerating disc degeneration, as shown in our finite element study. Stabilizing the segment with an M-PEEK rod may limit excessive motion, providing stability and maintaining intradiscal pressure closer to that of an intact model.