错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The effect of corneal power on the accuracy of 14 IOL power formulas

  • Jialin Xu,
  • Lu Zhang,
  • Er Mo,
  • Kaiyi Zhu,
  • Yitong Zhu,
  • Ke Feng,
  • Zunting Wu,
  • Yangran Zheng,
  • Fang Huang,
  • Xianhui Gong,
  • Jin Li

摘要

Background

This study evaluates the impact of corneal power on the accuracy of 14 newer intraocular lens (IOL) calculation formulas in cataract surgery. The aim is to assess how these formulas perform across different corneal curvature ranges, thereby guiding more precise IOL selection.

Methods

In this retrospective case series, 336 eyes from 336 patients who underwent cataract surgery were studied. The cohort was divided into three groups according to preoperative corneal power. Key metrics analyzed included mean prediction error (PE), standard deviation of PE (SD), mean absolute prediction error (MAE), median absolute error (MedAE), and the percentage of eyes with PE within ± 0.25 D, 0.50 D, ± 0.75 D, ± 1.00 D and ± 2.00 D.

Results

In the flat K group (Km < 43 D), VRF-G, Emmetropia Verifying Optical Version 2.0 (EVO2.0), Kane, and Hoffer QST demonstrated lower SDs (± 0.373D, ± 0.379D, ± 0.380D, ± 0.418D, respectively) compared to the VRF formula (all P < 0.05). EVO2.0 and K6 showed significantly different SDs compared to Barrett Universal II (BUII) (all P < 0.02). In the medium K group (43 D ≤ Km < 46 D), VRF-G, BUII, Karmona, K6, EVO2.0, Kane, and Pearl-DGS recorded lower MAEs (0.307D to 0.320D) than Olsen (OLCR) and Castrop (all P < 0.03), with RBF3.0 having the second lowest MAE (0.309D), significantly lower than VRF and Olsen (OLCR) (all P < 0.05). In the steep K group (Km ≥ 46D), RBF3.0, K6, and Kane achieved significantly lower MAEs (0.279D, 0.290D, 0.291D, respectively) than Castrop (all P < 0.001).

Conclusions

The study highlights the varying accuracy of newer IOL formulas based on corneal power. VRF-G, EVO2.0, Kane, K6, and Hoffer QST are highly accurate for flat corneas, while VRF-G, RBF3.0, BUII, Karmona, K6, EVO2.0, Kane, and Pearl-DGS are recommended for medium K corneas. In steep corneas, RBF3.0, K6, and Kane show superior performance.