Background <p>Computed Tomography (CT) plays a crucial role for diagnostic purpose; however, delivering optimal radiation dose to patients is the common challenge.</p> Objective <p>The main objective of this study was to establish abdomino-pelvic national diagnostic reference levels (NDRL) in Ethiopia.</p> Methods <p>A quantitative cross-sectional study design was employed in 23 health facilities. A total of 800 abdomino-pelvic CT scans of patients were evaluated from January, 2023 to March, 2024. Scan parameters, patient profile and CT dose describers were collected using a structured data collection format. Third quartile median values of the volumetric CT dose index (CTDI<sub>vol</sub>) and total dose length product (TDLP) were calculated using Microsoft Excel 2016 and SPSS software version 26. Finally, the results were compared with national, regional, and international DRL.</p> Result <p>The third quartile values obtained from the median values of TDLP and CTDI<sub>vol</sub> that were used as NDRL were 1387 mGy.cm and 10.5 mGy, respectively. This study found the highest and lowest median TDLP values of 3370mGy.cm and 273.7mGy.cm respectively.</p> Conclusion <p>The authors of this manuscript recommend that the NDRLs presented in this document can be used as a baseline against which hospitals future CT abdomino-pelvic median dose describers in Ethiopia can be compared.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Recommended National diagnostic reference levels for adult abdomino-pelvic computed tomography examinations in Ethiopia

  • Solomon Getachew Mekonnen,
  • Seife Teferi Dellie,
  • Teklehaimanot Mezgebe Nguse,
  • Daniel Zewdneh Solomon

摘要

Background

Computed Tomography (CT) plays a crucial role for diagnostic purpose; however, delivering optimal radiation dose to patients is the common challenge.

Objective

The main objective of this study was to establish abdomino-pelvic national diagnostic reference levels (NDRL) in Ethiopia.

Methods

A quantitative cross-sectional study design was employed in 23 health facilities. A total of 800 abdomino-pelvic CT scans of patients were evaluated from January, 2023 to March, 2024. Scan parameters, patient profile and CT dose describers were collected using a structured data collection format. Third quartile median values of the volumetric CT dose index (CTDIvol) and total dose length product (TDLP) were calculated using Microsoft Excel 2016 and SPSS software version 26. Finally, the results were compared with national, regional, and international DRL.

Result

The third quartile values obtained from the median values of TDLP and CTDIvol that were used as NDRL were 1387 mGy.cm and 10.5 mGy, respectively. This study found the highest and lowest median TDLP values of 3370mGy.cm and 273.7mGy.cm respectively.

Conclusion

The authors of this manuscript recommend that the NDRLs presented in this document can be used as a baseline against which hospitals future CT abdomino-pelvic median dose describers in Ethiopia can be compared.