Performance evaluation of a point of care capillary blood test for C-reactive protein in community healthcare
摘要
Antibiotics are often prescribed without a C-Reactive Protein (CRP) level, an inflammatory marker used to guide treatment decisions, due to delays in obtaining laboratory results. We evaluated a novel point of care testing (POCT) device that measures CRP within minutes using a small capillary blood sample obtained via finger prick, in community healthcare settings.
MethodPatients in the community having blood samples taken for CRP measurement provided finger prick capillary blood in children and venous blood in adults, who also provided an additional capillary sample. All capillary samples were tested using the LumiraDX™ POCT device, which provided a quantitative CRP measurement within four minutes. Venous samples from adults and the same capillary samples from children were then tested using a standard laboratory method.
ResultsA total of 161 patients agreed to measurements by the two methods. CRP levels as assessed by the laboratory method ranged from 0.9 to 382 mg/L, and 5.3 to 232.1 mg/L by POCT. There was a good correlation between results of the standard laboratory method and the POCT assay for CRP (r = 0.97), with a mean bias of -11.3 (slope 0.73, intercept 3.9 mg/L). Of the 161 patients tested, 57 gave CRP results outside the limits of quantitation of the POCT assay and were not included in the regression and bias analysis.
ConclusionThe LumiraDX™ POCT device demonstrates sufficient accuracy in measuring CRP levels from finger prick capillary blood. The availability of this technology across community healthcare settings will improve the assessment of infection and support more precise antibiotic prescribing.
Trial registrationACCom/SE/202,324/01 (Date of registration 17/08/2023).