Comparison of the oropharyngeal leak pressure in head rotation between a first-generation laryngeal mask and the i-gel laryngeal mask during general anesthesia - a randomized controlled trial
摘要
The laryngeal mask airway (LMA) is essential in anesthesia, using an air-inflatable cuff for optimal pharyngeal fit. The i-gel® LMA (i-gel) is unique for its thermoplastic elastomer material making an air-inflatable cuff unnecessary and has shown higher oropharyngeal leak pressure (OLP) than first-generation LMA (cLMA) in neutral head position, however, mixed results for head rotation. This prospective study aimed to compare OLP between the cLMA and i-gel during head rotation.
MethodsPatients undergoing elective surgery securing the airway with either cLMA or i-gel were included. Exclusion criteria included a Body Mass Index > 30 kg/m², cervical spine pathologies or restricted head rotation. OLP was measured after induction at 60° bilateral head rotation for primary outcome. Secondary outcomes were OLP at 0° head rotation and postoperative complications on postoperative day 1.
Results116 patients were randomized to either the cLMA (n = 60) or i-gel (n = 56) group. No significant difference was found in the OLP with 60° head rotation to the left (cLMA 20 cmH2O [Interquartile range [IQR] 18–22], i-gel 21 cmH2O [IQR 18-22.75], P = 0.752) and the right (cLMA 20 cmH2O [IQR 18–22], i-gel 20 cmH2O [IQR 15–21], P = 0.324). OLP was also comparable in a neutral head position (P = 0.368). No significant differences in postoperative outcome were observed.
ConclusionsThis study demonstrated comparable OLP between cLMA and i-gel during head rotation. This indicates equivalent efficacy and seal integrity of both LMAs during head rotation, resulting in sufficient tidal volumes during rotated head positions.
Trial registrationGerman Clinical Trials Register, retrospectively registered under DRKS00031785 (date of registration 26.04.2023).