Simultaneous Determination of Acetaminophen and Uric Acid in Biological Samples Using Differential Pulse Voltammetry and Orthogonal Signal Correction-Partial Least Squares
摘要
Abstract
Simultaneous determination of acetaminophen and uric acid in biological samples was performed using differential pulse voltammetry (DPV) on a glassy carbon electrode coupled with orthogonal signal correction-partial least squares (OSC-PLS). The optimum parameters were set at pH 8.0, a scan rate of 10 mV/s, and a pulse height of 50 mV. Both analytes showed peaks at about 390 mV. OSC-PLS was performed to solve the interference problem in the voltammograms. DPV calibration showed broad dynamic ranges of 10.0–220.0 and 10.0–210.0 µmol/L for acetaminophen and uric acid, respectively. The method was successfully applied to human plasma and urine samples without any pretreatment.