Abstract <p>Imaging Plates (IP) are widely used in laser-plasma and radiographic experiments for detection of electrons, ions, and X-rays. The use of IP in plasma diagnostics allows for obtaining quantitative information about plasma parameters. This paper presents the results of calibration of FujiFilm Imaging Plates BAS-MS and BAS-TR for X-rays in the range up to 10 MeV. Calibration was performed using the Ge-68 radioactive positron source and the Durr Dental VistaScan Mini View laser medical scanner.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Calibration of Imaging Plates for Detecting X-Rays

  • A. O. Khurchiev,
  • A. V. Skobliakov,
  • V. A. Volkov,
  • M. A. Lapshin,
  • S. A. Vysotskii,
  • A. V. Kantsyrev,
  • A. A. Golubev

摘要

Abstract

Imaging Plates (IP) are widely used in laser-plasma and radiographic experiments for detection of electrons, ions, and X-rays. The use of IP in plasma diagnostics allows for obtaining quantitative information about plasma parameters. This paper presents the results of calibration of FujiFilm Imaging Plates BAS-MS and BAS-TR for X-rays in the range up to 10 MeV. Calibration was performed using the Ge-68 radioactive positron source and the Durr Dental VistaScan Mini View laser medical scanner.