Evaluation of the Response Function of NaI (Tl) and Plastic Scintillator with Temperature Fluctuations
摘要
The temperature effect plays an important role in the response function and efficiency of detectors. In this study, the behavior of doped NaI (Tl) and plastic scintillators was evaluated and compared with temperature changes, in a similar condition, for the first time. The detection system uses an experimental setup consisting of a 60Co source, NaI (Tl) detector, and a plastic scintillation detector. The results show that the recorded count rates from NaI (Tl) detector have a direct relation with temperature and lead to a 0.15%/°C error in the count rate. But in the same condition, the acquired count rates from the plastic scintillator detector were reduced with temperature increasing with a 0.57%/°C error in the count rate (inverse relation). Also, in this work, Linear and second-order polynomial regression were implemented on the recorded count rates from scintillators to compensate temperature effect. Finally, the acquired results were evaluated using the relative error (%) and diagram slope which expressed the superiority of the polynomial regression. Using this correction technique, the count rate changes reached to least and are stable with temperature fluctuations. The results of this research can be suitable for the industrial application of NaI (Tl) and plastic scintillators in level gauging, thickness gauging, and other nuclear gauging systems.