Investigating the impact of OCT imaging of the crystalline lens on the accuracy and precision of cataract assessment

  • Niamh Burke,
  • Pádraig J. Mulholland,
  • Pearse A. Keane,
  • Julie-Anne Little

摘要

Purpose

To determine if supplementing standard clinical assessments with Optical Coherence Tomography (OCT) imaging of the crystalline lens improves the accuracy and precision of lens opacity assessment and associated clinical management decisions by optometrists.

Methods

Fifty optometrists registered in the UK or Éire undertook a clinical vignette study where participants graded lens opacities and made associated clinical management decisions based on the image(s)/information displayed. Three forms of vignettes were presented: (1) Slit-lamp (SL) images of the lens, (2) SL and OCT images and (3) SL, OCT and visual function measures. Vignettes were constructed using anonymised data from 50 patients with varying cataract severity, each vignette being presented twice in a randomised order (total vignette presentations = 300). The accuracy of opacity and management decisions were evaluated using descriptive statistics and non-parametric Bland–Altman analysis where assessments from experienced clinicians were the reference. The precision of assessments was examined for each vignette form using non-parametric Bland–Altman analysis.

Results

All (n = 50) participants completed the study, with 36 working in primary eyecare (primary eyecare) settings and 14 in hospital eyecare services (HES). Agreement was highest where vignettes contained all clinical data (i.e., SL, OCT and visual function data—grading: 51.0%, management: 50.5%), and systematically reduced with decreasing vignette content (p < 0.001). A larger number of vignettes containing imaging and visual function measures exhibited below reference (i.e., less conservative) grading compared with vignettes containing imaging data alone (all p < 0.05). HES-based optometrists were more likely to grade lens opacities lower than clinicians working in primary eyecare (p < 0.001). Good measurement precision was evident for all vignettes, with a mean bias close to zero and limits of agreement below one grading step for all conditions.

Conclusions

The addition of anterior segment OCT to SL images improved the accuracy of lens opacity grading. Structural assessment alone yielded more conservative decision making, which reversed once visual functional data was available.