<p>Three-dimensional (3D) image reconstructions are often rendered as two-dimensional images, using maximum intensity projections (MIPs). However, MIP’s rendering fidelity depends on the alignment of the individual slices along the projection direction. Also, the presence of noise and artifacts affects the contrast and the projected image elements. We introduce enhanced MIP (eMIP), a methodology that aligns the boundaries (e.g., skin boundary) of adjacent slices of the 3D volume onto the same coordinate system assumed by MIP (e.g., same depth) and applies robust contrast adjustment to normalize the intensities of the projected slices. We benchmark eMIP on 1725 clinical scans of human skin, using raster-scan optoacoustic mesoscopy (RSOM) that were assessed by 8 experts. Our results show that eMIP facilitates interpretability compared to conventional MIP and increases consistently the perceived image quality. The improved diagnostic ability of eMIP has the potential to replace MIP in RSOM and similar modalities.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Enhanced maximum intensity projection (eMIP) for improving the fidelity of optoacoustic images

  • Manuel Gehmeyr,
  • María Begoña Rojas López,
  • Suhanyaa Nitkunanantharajah,
  • Hubert Preißl,
  • Andreas Vosseler,
  • Reiner Jumpertz von Schwartzenberg,
  • Andreas L. Birkenfeld,
  • Nikoletta Katsouli,
  • Nikolina-Alexia Fasoula,
  • Angelos Karlas,
  • Michael Kallmayer,
  • Anette-Gabriele Ziegler,
  • Dominik Jüstel,
  • Vasilis Ntziachristos

摘要

Three-dimensional (3D) image reconstructions are often rendered as two-dimensional images, using maximum intensity projections (MIPs). However, MIP’s rendering fidelity depends on the alignment of the individual slices along the projection direction. Also, the presence of noise and artifacts affects the contrast and the projected image elements. We introduce enhanced MIP (eMIP), a methodology that aligns the boundaries (e.g., skin boundary) of adjacent slices of the 3D volume onto the same coordinate system assumed by MIP (e.g., same depth) and applies robust contrast adjustment to normalize the intensities of the projected slices. We benchmark eMIP on 1725 clinical scans of human skin, using raster-scan optoacoustic mesoscopy (RSOM) that were assessed by 8 experts. Our results show that eMIP facilitates interpretability compared to conventional MIP and increases consistently the perceived image quality. The improved diagnostic ability of eMIP has the potential to replace MIP in RSOM and similar modalities.