<p>Photonic ultrasound sensors promise unparalleled spatial and temporal resolution in ultrasound imaging due to their size-independent noise figure, high sensitivity, and broad bandwidth. Optical materials can further improve performance and stability, but achieving small size, high sensitivity, and wide bandwidth remains challenging. This work introduces amorphous silicon carbide (a-SiC) for ultrasound sensing, offering strong optical confinement, low propagation loss, and high stability for miniaturized microring sensors. We demonstrate a compact detection system with a 20-transducers linear array coupled to a single bus waveguide. The sensors achieve an optical finesse of 1320 and intrinsic sensitivity of 78 fm kPa<sup>−1</sup>, leading to a noise-equivalent pressure below 55<InlineEquation ID="IEq1"> <EquationSource Format="TEX">\({{{\rm{mPa}}}}/\sqrt{{{{\rm{Hz}}}}}\)</EquationSource> <EquationSource Format="MATHML"><math> <mi mathvariant="normal">mPa</mi> <mo>/</mo> <msqrt> <mrow> <mi mathvariant="normal">Hz</mi> </mrow> </msqrt> </math></EquationSource> </InlineEquation>, calibrated from 3.36 MHz to 30 MHz. High-resolution imaging of fine structures validates real-world applicability. a-SiC is also easily integrated on most substrates due to its low deposition temperature. Our results position a-SiC as a promising solution for optical ultrasound sensing, combining miniaturization, low-loss, and high-sensitivity.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Amorphous silicon-carbide photonics for ultrasound imaging

  • R. Tufan Erdogan,
  • Bruno Lopez-Rodriguez,
  • Wouter J. Westerveld,
  • Sophinese Iskander-Rizk,
  • Gerard J. Verbiest,
  • Iman Esmaeil Zadeh,
  • Peter G. Steeneken

摘要

Photonic ultrasound sensors promise unparalleled spatial and temporal resolution in ultrasound imaging due to their size-independent noise figure, high sensitivity, and broad bandwidth. Optical materials can further improve performance and stability, but achieving small size, high sensitivity, and wide bandwidth remains challenging. This work introduces amorphous silicon carbide (a-SiC) for ultrasound sensing, offering strong optical confinement, low propagation loss, and high stability for miniaturized microring sensors. We demonstrate a compact detection system with a 20-transducers linear array coupled to a single bus waveguide. The sensors achieve an optical finesse of 1320 and intrinsic sensitivity of 78 fm kPa−1, leading to a noise-equivalent pressure below 55 \({{{\rm{mPa}}}}/\sqrt{{{{\rm{Hz}}}}}\) mPa / Hz , calibrated from 3.36 MHz to 30 MHz. High-resolution imaging of fine structures validates real-world applicability. a-SiC is also easily integrated on most substrates due to its low deposition temperature. Our results position a-SiC as a promising solution for optical ultrasound sensing, combining miniaturization, low-loss, and high-sensitivity.