<p>Respiration is a key indicator of health and wellbeing, with metrics such as respiratory rate (RR), inspiratory time (TI), expiratory time (TE), and the inspiratory-to-expiratory time (I:E) ratio offering insights into conditions ranging from acute life-threatening and chronic diseases to symptom management. While traditional methods already measure these parameters with high accuracy, they still require contact-based sensors, limiting their practicality for continuous monitoring. This study evaluates radar as a non-contact alternative by validating multiple radar-derived respiratory metrics against impedance pneumography measurements in 30 healthy volunteers at rest. Synchronous recordings from both modalities were analysed to assess agreement across methods using descriptive statistics, scatter plots, modified Bland-Altman plots, and equivalence testing (TI: ±0.3&#xa0;s, TE: ±0.3&#xa0;s, RR: ±2 brpm, I:E ratio: ±0.2). Equivalence testing indicated high correlation (<i>p</i> ≤ 0.001***) across all metrics, with 81.8% (TI), 77.6% (TE), 97.2% (RR), and 85.7% (I:E ratio) of values within predefined bounds. These findings highlight radar’s potential for continuous respiratory monitoring, particularly in medical fields where minimizing patient burden is essential as in palliative, post anaesthesia, and intensive care settings.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Radar-based inspiratory-to-expiratory time ratio estimation: a validation study

  • Thanh Trúc Trần,
  • Marie Oesten,
  • Stefan G. Griesshammer,
  • Anke Malessa,
  • Kilin Shi,
  • Maria Heckel,
  • Bjoern M. Eskofier,
  • Alexander Koelpin,
  • Christoph Ostgathe,
  • Tobias Steigleder

摘要

Respiration is a key indicator of health and wellbeing, with metrics such as respiratory rate (RR), inspiratory time (TI), expiratory time (TE), and the inspiratory-to-expiratory time (I:E) ratio offering insights into conditions ranging from acute life-threatening and chronic diseases to symptom management. While traditional methods already measure these parameters with high accuracy, they still require contact-based sensors, limiting their practicality for continuous monitoring. This study evaluates radar as a non-contact alternative by validating multiple radar-derived respiratory metrics against impedance pneumography measurements in 30 healthy volunteers at rest. Synchronous recordings from both modalities were analysed to assess agreement across methods using descriptive statistics, scatter plots, modified Bland-Altman plots, and equivalence testing (TI: ±0.3 s, TE: ±0.3 s, RR: ±2 brpm, I:E ratio: ±0.2). Equivalence testing indicated high correlation (p ≤ 0.001***) across all metrics, with 81.8% (TI), 77.6% (TE), 97.2% (RR), and 85.7% (I:E ratio) of values within predefined bounds. These findings highlight radar’s potential for continuous respiratory monitoring, particularly in medical fields where minimizing patient burden is essential as in palliative, post anaesthesia, and intensive care settings.