Image quality evaluation of imaging skins, a novel stretchable X-ray detector for intraoperative tumour imaging
摘要
This paper presents the development of a novel X-ray detector composed of silicone elastomer and GOS:Tb, which we refer to as imaging skins. These detectors were integrated into a custom X-ray system to convert radiation into visible light. Our study focused on how fabrication parameters such as thickness and concentration impact sensor linearity, considering their potential application directly on the skin or organs to identify tumour margins during surgery. In addition, we examined how the stretching capabilities of these detectors influenced the image quality. Our imaging detection stack demonstrated consistent linearity across various fabrication parameters with the coefficient of determination (