错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Continuous spatial field confocal thermometry using lanthanide doped tellurite glass

  • Daniel Stavrevski,
  • E. P. Schartner,
  • Q. Sun,
  • I. S. Maksymov,
  • R. A. McLaughlin,
  • H. Ebendorff-Heidepriem,
  • A. D. Greentree

摘要

Distinguishing between microscopic variances in temperature in both space and time with high precision can open up new opportunities in optical sensing. In this paper, we present a novel approach to optically measure temperature from the fluorescence of erbium:ytterbium doped tellurite glass, with fast temporal resolution at micron-scale localisation over an area with sub millimetre spatial dimensions. This confocal-based approach provides a micron-scale image of temperature variations over a 200 \(\upmu \) μ m \(\times \) × 200 \(\upmu \) μ m field of view at sub-1 second time intervals. We test our sensing platform by monitoring the real-time evaporation of a water droplet over a wide field of view and track it’s evaporative cooling effect on the glass where we report a net temperature change of 6.97 K ± 0.03 K. This result showcases a confocal approach to thermometry to provide high temporal and spatial resolution over a microscopic field of view with the goal of providing real-time measures of temperature on the micro-scale.